25
CAL
INTV
SERIAL COMMUNICATION COMMAND PREFIX SET
Press
CAL
when the command prefix is displayed. The command prefix (0 to
47) will start blinking.
Press the
ARROW
keys to change the command prefix.
Press
CFM
to save the modified command prefix value.
Press
CAL
to escape without saving.
Note:
• See the PC interface section on page 29 for a complete explanation.
• The command prefix must be 16 if
HI 92000
PC software is used.
CALIBRATION ALARM TIME-OUT SET
Press
CAL
when the calibration time-out is displayed. Calibration time-out
(OFF or 1 to 7 days) will start blinking.
Press the
ARROW
keys to change the calibration time-out value.
Press
CFM
to save the modified calibration time-out value.
Press
CAL
to escape without saving.
INSTRUMENT ID PARAMETER SET
Press
CAL
when the instrument ID is displayed. The instrument ID (0000 to
9999) will start blinking.
Press the
ARROW
keys to change the instrument ID value.
Press
CFM
to save the modified instrument ID value.
Press
CAL
to escape without saving.
Note:
The instrument ID is downloaded to a PC as part of a logged data, set
to identify its origin.