47
• Press CFM to confirm the parameter that stops blinking
on the primary display.
The secondary display shows the multimeter or HI931002
input value as the interval lower limit.
• Use
or
to make the HI931002
or multimeter output correspond with
the value shown on the secondary
display (e.g. 4).
• Wait until the calibrator reading is stable (approx.30 sec.)
• Press CFM to enter. The meter will switch to the second
calibration point. Repeat the above procedure.
• After the desired readings are obtained, press CFM and
the meter will skip back to normal operating mode.
Note
When adjusting values using the
or
it is important to
allow for sufficient response time (up to 30 seconds)
The table below lists the values of output codes along with
the calibration values (which are the analog output minimum
and maximum) as indicated on the display.
The secondary display indicates the current calibration
value, while primary display indicates the current calibra-
tion type.
OUTPUT TYPE
CALIBRATION CODE CALIBRATION POINT 1 CALIBRATION POINT 2
0-1 mA
0
0 mA
1 mA
0-20 mA
1
0 mA
20 mA
4-20 mA
2
4 mA
20 mA
0-5 Vdc
3
0 Vdc
5 Vdc
1-5 Vdc
4
1 Vdc
5 Vdc
0-10 Vdc
5
0 Vdc
10 Vdc