21
21
Select the Manual (Stability) lot and press
SETUP/CLR
key. The
”CLEAR” message will be displayed along with ”MANUAL” or
“STABILITY” and “CFM” tag blinking on the LCD. Press the
GLP/CFM
key to confirm the deleting of the selected lot (MANUAL or STABILITY)
or all records. Press
SETUP/CLR
key to exit without deleting.
The lot number is used to identify particular sets of data. The lot numbers are allocated successively
until 100, even if some lots were deleted. The total number of lots that can be saved is 100. If
some are deleted (for example 1‑50), fifty additional logs may be stored. These will be numbered
101‑150. The lots are allocated successively (provided available memory space) until 999 is reached.
After this, it is necessary to delete all the LOT logs to start over the numbering.
Delete All
All pH logs, (or all ORP logs) may be deleted in a single clear. This function will delete all MANUAL,
STABILITY and INTERVAL logs.
Press the
RCL
key. pH will be blinking. Use
RANGE
/
u
key
to select
desired measurement parameter log data to delete (pH or ORP).
While the measurement type is blinking and message states “LOG
RECALL”, press
SETUP/CLR
key.
“CLEAR ALL” and measurement type will be displayed with “CFM” tag
blinking. Press
GLP/CFM
key.
“PLEASE WAIT” and the percent cleared will be displayed until
completed.
Note: If SETUP/CLR key is pressed in error, press SETUP/CLR
key again to exit without deleting.
SETUP/ INST
ALLA
TION