Stability Logging
Select Stability and choose measurement stability criteria in the
SETUP menu. Only Stability Medium is available in Basic mode. To
initiate the Stability log, press the
LOG
key while the instrument is
in measurement.
The “PLEASE WAIT” screen will be displayed briefly followed by a
screen showing the stability tag, “LOG” tag and a “WAITING”
message. The log can be stopped while the “WAITING” message is
displayed by pressing
LOG
key again.
When the stability selected criteria has been met, a “SAVED” message
will be displayed followed by a screen indicating how much log space
is available. The “LOG” tag will be displayed on all 4 screens.
“PLEASE WAIT”
“WAITING”
“SAVED” with the log record number
“FREE” with the number of free spaces available
All log records stored on
edge
® pH
. may be viewed on the meter by pressing the
RCL
key. The log
records are grouped according to measurement (pH, ORP).
The parameter displayed first is based on the probe/sensor connected.
The display also indicates the percentage of log memory used. Press
GLP/CFM
key to display those logs.
Choices are:
• Manual log on demand lot,
• Manual log on stability lot,
• Individual Interval logging lots.
If no data was logged for the selected measurement range, the
instrument displays the following messages (ex. for pH range):
“NO MANUAL LOGS”
“NO STABILITY LOGS”
VIEWING LOGGED DATA
SETUP/ INST
ALLA
TION