TIONS
3. SPECIFICATIONS
pH Specifications
EC Specifications
TDS Specifications
*Note: 1000 μS/cm = 500 ppm with 0.5 TDS Factor
= 700 ppm with 0.7 TDS Factor
Range
0.00 to 14.00 pH, 0.0 to 14.0 pH
Resolution
0.01 pH, 0.1 pH
Accuracy (@25 ºC/77 ºF)
±0.05 pH, ±0.1pH
Calibration
one or two‑point calibration (using pH 4.01,7.01, 10.01
buffers) using auto buffer recognition.
one‑point calibration using quick calibration solution
Temperature Compensation
Automatic: 0.0 to 60.0 ºC; 32.0 to 140.0 °F
Range
0.00 to 10.00 mS/cm
Resolution
0.01 mS/cm
Accuracy (@25 ºC/77 ºF)
±0.1mS/cm (0.00 to 5.00mS/cm)
±0.2mS/cm (5.00 to 10.00mS/cm)
Calibration
one‑point at 1.41 mS/cm or 5.00 mS/cm using auto
standard recognition.
one‑point calibration using quick calibration solution
Temperature Compensation
(not user selectable)
automatic, with
b
= 1.9%/°C
Range
0 to 5000 ppm (0.5 TDS Factor)*
0 to 7000 ppm (0.7 TDS Factor)*
Resolution
10 ppm
Accuracy (@25 ºC/77 ºF)
±2%FS
Calibration
through EC calibration
TDS Factor
0.5(500 ppm) or 0.7(700 ppm)