6.2. GLP INFORMATION
Good Laboratory Practice (GLP) refers to a quality control function used to ensure uniformity and
consistency of sensor calibrations and measurements. Select
pH
or
EC
from MENU and press
GLP
to review the calibration information. The monitor can store calibration info from the last 5 pH and
EC calibrations. The newest calibration is positioned at #1. Pressing
▲▼
keys in GLP can view the
informations from last five calibrations.
GLP pH
Highlight
pH
from
MENU
and press
GLP
. Monitor will display date/time, buffer used, slope and
offset from last 5 calibrations.
If a single point buffer calibration was made, the previous calibration slope will be used.
If all calibrations were cleared “ No user calibration” message is displayed.
GLP EC
Highlight EC from
MENU
and press
GLP
. The monitor will display date/time, standard and cell
coefficient from the last 5 calibrations. The cell coefficient can be used to determine when the probe
requires cleaning. If this number decreases below 75% the probe requires cleaning.
If all calibrations were cleared “ No user calibration” message is displayed