Connecting the Probe to the Pump Controller (Flow Cell Configuration)
• Remove the protective cap and verify if the O-ring (2) is in place.
• The probe should be connected to the controller and calibrated before installation.
• To avoid twisting the cable, unplug probe from socket temporarily while installing in flow cell.
• Carefully insert the probe (1) into the flow cell adapter (3), paying attention not to damage the
O-ring. Tighten the probe enough to seal in place.
• Cut the tubing (7) to reach between the inlet valve mounted on the pipe and the flow cell inlet.
Push the end of the tubing (7) into the elbow tubing fitting.
• Repeat the previous step for flow cell outlet pushing the tubing (5) into the elbow tubing fitting (4).
Position Description
1
Probe
2
O-ring
3
Flow cell adapter
4
Elbow tubing fitting
5
Tubing
6
Elbow tubing fitting
7
Tubing
Note:
Prepare and calibrate the probe prior to installing in the flow cell.