25
SETUP
High Alarm Value
Option:
user selectable
The range is influenced by the low-alarm value set (e.g. if low-alarm value is set as 5 pH, high-alarm
value can be set from 5.10 pH and incremented up to 14.00 pH).
“HIGH ALARM VALUE” message is scrolled on the bottom of the LCD screen.
Note:
The user can set the high-alarm value with high-alarm option enabled only.
Low Alarm
Option:
Enabled (En) or disabled (diS)
Press the arrow keys to switch between the options.
“LOW ALARM” message is scrolled on the bottom of the LCD screen.
Low Alarm Value
Option:
user selectable
The range is influenced by the high-alarm value set (e.g. if high-alarm value is set as 8 pH, low-alarm
value can be set from 0.00 pH and incremented up to 7.90 pH).
“LOW ALARM VALUE” message is scrolled on the bottom of the LCD screen.
Note:
The user can set the low-alarm value with low-alarm option enabled only.
Temperature Unit
Option:
user selectable: °C or °F
Press the arrow keys to change the unit.
“TEMPERATURE UNIT” message is scrolled on the bottom of the LCD screen.