Setup/Installation

14

In Setup mode, choose log parameter, press 

MODIFY 

then use the 

u

 arrow to 

select between Interval, Manual, or Stability. When Interval is displayed, use the 

p

 and 

q

 arrows to select the setting for the timed interval. When Stability is 

displayed, use the 

p

 and 

q

 to select the measurement stability setting.

A  complete  set  of  GLP  information  including  date,  time,  range  selection, 
temperature reading, calibration information and probe serial number is stored 
with each log made.

Interval Logging

Select Interval and sampling period in the SETUP menu (Not available in Basic 
mode). To start Interval logging, press the 

LOG 

 key while the instrument is in 

measurement mode.

A  “PLEASE WAIT” message will be displayed followed 
by  the  number  of  free  spaces.  During  active  interval 
logging,  lot  information  is  displayed  on  the  third 
LCD  line.  The  line  indicates  in  which  lot  the  data  will 
be  placed  and  keeps  track  of  the  number  of  logged 
records taken. The “LOG“ tag is continuously on during 
active logging. 

Pressing 

u

  during  an  interval  log  will  display  the 

number of logs available.

Pressing  the 

LOG 

  key  again  will  stop  the  Interval 

logging session. The “LOG STOPPED“ message will be 
displayed for a few seconds.

If a sensor failure occurs during interval logging, the  
message  “OUT  OF  SPEC.”  will  alternate  with  logging 
information.

Manual Logging

Select  Manual  in  the  SETUP  menu.  To  initiate  a 
Manual log, press the 

LOG 

 key while the instrument 

is  in  measurement.  The  “PLEASE  WAIT”  screen  will 
be  displayed  briefl y  followed  by  a  screen  indicating 
the measurement has been saved and then a screen 
indicating the log record number. 

The “LOG” tag will be displayed on all 3 screens.
“PLEASE WAIT”
“SAVED” with the log record number
“FREE” with the number of free spaces available

Logging 

Function

Содержание edge EC HI2003

Страница 1: ...EC ...

Страница 2: ...nsent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Included 3 Safety Measures 3 Description 4 Diagram 4 8 Product Diagram 4 5 Probe Diagram 6 Keypad Function 7 Guide to Indicators 8 Setup Installation 9 19 Setting Up edgeEC 9 10 Electrode Probe Connections 10 General Setup 11 12 Basic Mode 13 Logging Function 13 15 Viewing Logged Data 15 17 PC Storage Interface 18...

Страница 3: ...ng this product make sure that it is entirely suitable for your specific application and for the environment in which it is used Operation of this instrument may cause interference to other electronic equipment requiring the operator to take steps to correct interference Any variation introduced by the user to the supplied equipment may degrade the instrument s EMC performance To avoid damages or ...

Страница 4: ...o offers a basic operation mode that streamlines measurement configuration and is useful for many routine applications Every featureandmeasurementdetailisdesignedtogiveyouanedgeinmeasurement technology edgeEC is versatile in many ways The slim meter and probe can be used as a portable device using its rechargeable battery or used in its bench or wall cradles that also power the meter as a line pow...

Страница 5: ... 2 Capacitive Touch Keypad 3 3 mm jack input for edgeEC digital probes 4 Top mounted ON OFF button 5 Micro USB device connection for power or PC interface 6 Standard USB host connection for data transfer to a USB thumb drive Product Diagram ...

Страница 6: ...o sensor recognition Store calibration specific data from the last calibration Are built with materials suitable for use in chemical analysis Have integrated temperature measurement Incorporate a 3 mm jack termination Unique serial ID in every probe for traceability Probe Probe Diagram ...

Страница 7: ...ETUP CLR Used to enter exit SETUP mode During calibration used to clear previous calibration data In log RCL used to clear log records 5 Used to scroll through SETUP menu Used to change selection when modifying a parameter in SETUP 6 RCL Recall Used to view logged records or view log memory used 7 LOG Used to log data by manual log on demand or manual log on stability or to start stop interval log...

Страница 8: ...ent 10 Temperature units 11 Temp Compensation status 12 Measurement line 13 Measurement units 14 Stability Indicator The third line of the LCD 9 is a dedicated message line During measurement the user may use the pq arrows to select desired message Options include date time calibration data battery charge or no message If a measurement error or log status change occurs during measurement the third...

Страница 9: ...ed for the measurement you will be making 6 Calibrate the sensor probe You are now ready for measurements Bench Cradle Setup Insert electrode holder arm into the post on the pivoting base Connect the probe connector to the socket located at the bottom of the instrument Slide edgeEC into the cradle while positioning the probe cable behind the cradle Put the probe sensor into the electrode holder an...

Страница 10: ...at the top Plug the 5 VDC adapter into the power supply socket or by connecting directly to a PC Note edgeEC is supplied with a rechargeable battery inside which provides about 8 hours of continuous use Whenever edgeEC is connected to the power adapter or to a PC the battery is charging Connect the 3 mm probe jack to the probe input located on the bottom of edgeEC Make sure the probe is completely...

Страница 11: ...G ON EDGE or EXPORT TO PC LOG ON EDGE Available Log Select log type to be used from 3 types of logging Manual log on demand Manual log on stability 3 types of stability criteria available Timed interval lot logging Manual Log Stability Log Fast Medium Accurate Interval Log Seconds 5 10 30 Interval Log Minutes 1 2 5 15 30 60 120 180 Interval 5 Sec Manual log or Stability log Medium Set Calibration ...

Страница 12: ...t Only seen when log errors are present Permits formatting the flash drive On or Off Off Available Message Transition User may choose how messages are displayed on third LCD line of display Word scroll messages or letter scroll messages Letter scroll messages Available Reset Config To Default Press the MODIFY key and CFM when prompted to reset parameters Available RESETS with Basic Mode OFF Instru...

Страница 13: ...is created The maximum number of Interval lots that may be stored is 100 If a 101st lot is attempted MAX LOTS will be displayed Some lots will need to be deleted The lot numbering is up to 999 and restarts if all lot logs are deleted All log on demand and stability log on demand are stored in a single lot The maximum number of records that may be stored in a Manual or Stability lot is 200 records ...

Страница 14: ...ing lot information is displayed on the third LCD line The line indicates in which lot the data will be placed and keeps track of the number of logged records taken The LOG tag is continuously on during active logging Pressing u during an interval log will display the number of logs available Pressing the LOG key again will stop the Interval logging session The LOG STOPPED message will be displaye...

Страница 15: ...SAVED with the log record number FREE with the number of free spaces available All log records stored on edgeEC may be viewed on the meter by pressing the RCL key If no sensor or probe is connected use the u arrow to select the desired measurement type Press CFM to display those logs Choices are Manual log on demand lot Manual log on stability lot Individual Interval logging lots If no data was lo...

Страница 16: ...og by pressing CFM when Manual Stability is displayed Use the ARROW keys to select the record to be deleted and then press CLR The instrument will display CLEAR REC and record numberalongwith CFM tagblinking UsetheARROW keys to select another record if necessary Press CFM The instrument will display PLEASE WAIT and then CLEAR DONE message When individual logs are deleted within saved MANUAL or STA...

Страница 17: ...y space until 999 is reached After this it is necessary to delete all the LOT logs to start over the numbering Delete All All logs may be deleted in a single clear This function will delete all MANUAL STABILITY and INTERVAL logs Press the RCL key while EC is displayed on the first LCD line and the third LCD line message states LOG RECALL Press CLR CLEAR ALL will be displayed with CFM and EC tags b...

Страница 18: ... appears followed by EXPORT Press CFM to export the selected record or lot If CFM is not pressed within 10 seconds the USB host will become inactive The meter will display the percentage of export The export percentage should go to 100 Remove USB flash drive If the selected file is already saved on the flash drive edgeEC will ask for confirmation of overwriting the existing file The message OVERWR...

Страница 19: ...set and English language are suggested settings Other files may be visible depending upon computer settings All files stored will appear in this folder Adjust Font or column width appropriately Adjust the decimal places All conductivity measurements will be displayed as μS cm Interval logs is designated as EC Lots ie ECLOT001 ECLOT002 ECLOT003 The Manual Lots are ECLOTMAN The Stability Lots are EC...

Страница 20: ...EC Mode Standard operation mode allows complete configuration of all options for the measurementofConductivity TDSorSalinitymeasurements Measurementdata can be logged using Manual log on demand Manual Log on Stability or Interval logs All logged data can be exported to a thumb drive or PC Basic operation mode is used for conductivity and TDS measurements Common default values are used for Cell Fac...

Страница 21: ...rect measured conductivity to a reference temperature by applying a fixed factor for linear compensation 0 00 to 6 00 ºC Note Setting to 0 00 is the same as using No TC 1 90 ºC Close for natural waters or salt solutions Not available Automatically set to 1 90 ºC T Ref ºC The user may select either 20ºC or 25ºC reference temperature for temperature corrected conductivity 20 ºC or 25 ºC 25 ºC Availa...

Страница 22: ...emain fixed throughout the measurement AUTO 14 99 mg L 149 9 mg L 1499 mg L 14 99 g L 100 0 g L 400 g L AUTO Not available but measurement autoranges as needed TDS Unit Select units of measurement for TDS measurement mg L or ppm ppm Available to select EC Salinity Scale Three measurement scales are available for salinity measurement in Seawater Practical Salinity Scale Percent Scale and Natural Se...

Страница 23: ...nstrument frequently especially if high accuracy is required The instrument should be recalibrated Whenever the EC probe is replaced At least once a week After testing aggressive chemicals If CAL DUE is displayed on the third LCD line Every time you calibrate the instrument use fresh standard and perform electrode maintenance as required It is recommended to choose a calibration standard that is c...

Страница 24: ...emperature coefficients during calibration ºC ºF HI7030 HI8030 μS cm HI7031 HI8031 μS cm HI7033 HI8033 μS cm HI7034 HI8034 μS cm HI7035 HI8035 μS cm HI7039 HI8039 μS cm 0 32 7150 776 64 48300 65400 2760 5 41 8220 896 65 53500 74100 3180 10 50 9330 1020 67 59600 83200 3615 15 59 10480 1147 68 65400 92500 4063 16 60 8 10720 1173 70 67200 94400 4155 17 62 6 10950 1199 71 68500 96300 4245 18 64 4 1119...

Страница 25: ...ap the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve Press CAL to enter calibration The STIR and CAL tags will turn on and the first LCD line will display the percent NaCl reading the second LCD line will display the temperature and the third LCD line will display WAIT message blinking until the reading is stable When the reading is stable and close to the standa...

Страница 26: ...To view more information enter GLP mode by pressing GLP key EC calibration data is stored automatically after a successful calibration To view the EC calibration data press GLP when the instrument is in EC measurement mode The instrument will display the calibration standard and the temperature of the calibrated standard Use the pq keys to scroll through the calibration data displayed on the third...

Страница 27: ...e current reading Calibration Expiration status together with the current reading If disabled EXPIRATION WARNING DISABLED is displayed If enabled the number of days until the calibration alarm CAL DUE will be displayed i e CAL EXPIRES IN 2 DAYS If enabled the number of days since the calibration expired i e CAL EXPIRED 2 DAYS AGO The serial number of the probe together with the current reading EC ...

Страница 28: ...he current reading The salinity coefficient determined from the calibration together with the current reading The Solution Standard together with the calibrated temperature The time hh mm ss of the last calibration together with the current reading The date yyyy mm dd of the last calibration together with the current reading Calibration Expiration status together with the current reading If disabl...

Страница 29: ...trapped inside the sleeve The conductivity value will be displayed on the first LCD line the temperature on the second LCD line and calibration or range specific information on the third LCD line To toggle between information displayed on the third LCD line use the ARROW keys If the reading is out of range when the range is set to automatic the full scale value 200 0 mS cm for ATC mode or 500 0 mS...

Страница 30: ... line The temperature will continue to be displayed blinking If temperature exceeds the meter specification of 120º C then 120º C will blink on the display If interval logging the message OUT OF SPEC will alternate with the Log specific messages in both of these cases The Log file will indicate a ºC next to the data In the case the temperature sensor is damaged BROKEN TEMPERATURE SENSOR along with...

Страница 31: ...ty units is calculated using the following formula NaCl Percent Scale The NaCl scale is an older salinity scale used for seawater salinity In this scale 100 salinity is equivalent to roughly 10 solids High percentages were made by evaporation To display NaCl in units enter SETUP and select NaCl unit Press the RANGE key until NaCl is displayed on the LCD The instrument will display the NaCl reading...

Страница 32: ... ratio and salinity is defined by the following equation S 0 08996 28 2929729R15 12 80832R15 2 10 67869R15 3 5 98624R15 4 1 32311R15 5 Note The formula can be applied for temperatures between 10 ºC and 31 ºC If the reading is out of the measurement range the display will flash the highest value possible and a warning message will be displayed If this occurs during a log a will be placed next to th...

Страница 33: ... clean the probe with a cloth or a nonabrasive detergent Make sure to reinsert the sleeve onto the probe properly and in the right direction After cleaning the probe recalibrate the instrument The insulator used to support the platinum rings is made of glass Use extreme caution when handling this probe Maintenance EC Probe Maintenance ...

Страница 34: ... Out of range in EC TDS or Salinity scale Verify the plastic shipping spacer has been removed from probe Recalibrate the probe Make sure the solution is in specified range Make sure the range is not locked Select Auto range The meter does not measure temperature is displayed on second LCD line Broken temperature sensor Replace the probe The meter fails to calibrate Salinity Incorrect EC calibratio...

Страница 35: ...bration 6 standards available 84 1413 μS cm 5 00 12 88 80 0 111 8 mS cm 1 point offset 0 00 μS cm Single point with HI 7037L Standard Temperature compensation Automatic 5 to 100º C 23 to 212º F NoTC none absolute conductivity Conductivity temperature coefficient 0 00 to 6 00 ºC for EC and TDS only Default value is 1 90 ºC TDS factor 0 40 to 0 80 default value is 0 50 EC Probe HI 763100 Log feature...

Страница 36: ...Specifications PC Interface Micro USB Storage Interface USB Power Supply 5 VDC Adapter included Environment 0 50 ºC 32 122 ºF Max 95 RH non condensing Dimensions 202 x 140 x 12 mm 7 9 x 5 5 z 0 5 Weight 250g 8 82 oz Specifications ...

Страница 37: ... cm 500 mL bottle HI 7034L 80000 μS cm 500 mL bottle HI 7035L 111800 μS cm 500 mL bottle HI 7039L 5000 μS cm 500 mL bottle HI 7037L 100 NaCl sea water standard solution 500 mL bottle HI 8030L 12880 μS cm 500 mL FDA approved bottle HI 8031L 1413 μS cm 500 mL FDA approved bottle HI 8033L 84 μS cm 500 mL FDA approved bottle HI 8034L 80000 μS cm 500 mL FDA approved bottle HI 8035L 111800 μS cm 500 mL ...

Страница 38: ...bed maintenance is not covered If service is required contact your local Hanna Office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization number from the Tech...

Страница 39: ...Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice MANEDGEEC 01 15 ...

Отзывы: