Thank you for choosing a Hanna Instruments product. Please read 
this instruction manual carefully before using the instrument. For 
additional information, visit www.hannainst.com or e-mail us at 
[email protected].

Range

-50.0 to 150.0 ºC / -58.0 to 302 ºF

Resolution

0.1 ºC (-50.0 to 150.0 ºC) 
0.1 ºF (-58.0 to 199.9 ºF) / 1 ºF (above 200 ºF)

Accuracy

±0.2 ºC (-30.0 to 120.0 ºC)
±0.3 ºC (outside: -50.0 to -30.0 ºC and 
120.0 to 150.0 ºC)
±0.5 ºF (-22.0 to 199.9 ºF)
±1 ºF (outside: -58.0 to -22.0 ºF
and 200 to 302 ºF)

Environment -30 to 50 ºC (-22 to 122 ºF)

Battery Type

CR2032 (1 pc.)

Battery Life

approx. 2000 hours of continuous use

Auto Off

8 min (default), 60 min. or OFF

Dimensions

50 x 185 x 21 mm (2 x 7.3 x 0.9“)

Weight

50 g (1.8 oz.)

Certification

EN 13485:2001 - Suitability: Storage;
Environment: E;  Accuracy Class: 0.5; IP65

Turn Checktemp

®

 ON with 

ON/OFF

 button: all LCD segments are 

displayed for 1 second, and CalCheck

®

 starts automatically to verify 

that Checktemp

®

 is still accurate to its specifications. A message of 

“CAL“

 and 

“-0-“

 means that Checktemp

®

 is within specifications and 

ready to be used. While measuring temperature, remember that:
•  The temperature sensor is located at the tip of the 

stainless steel probe.

•  It is necessary to wait for a few seconds for the display to stabilize.

Liquid Crystal Display

ON/OFF button

Maximum Immersion Level

Temperature Probe

4 cm 
1.5”

Recommended  
Immersion Level

EN

13

48

5

Sto

rag

e

E /

0.5

Battery
Setup 
button

Enter Setup to change the temperature unit (ºC or ºF) or to adjust the 

auto shut off time. To enter the Setup MODE, press the 

SET

 button 

located inside the battery compartment once for temperature unit or 

twice for auto shut off.

• Use the 

ON/OFF

 button to switch between “ºC” or “ºF”. Press 

SET

 

once to save and continue to auto shut off settings.

• Use the 

ON/OFF

 button to cycle through the following Auto Shut 

Off options: 8 minutes (“d08”, default value), 60 minutes (“d60”), 
OFF (“d--”). Press the 

SET

 button once to save and continue to 

normal operation.

Replace the battery when the battery symbol is blinking or when 

Checktemp

®

 does not switch on. If the meter has been submerged in 

a liquid, make sure it has been completely dried. Check the polarity 

of the battery, open the compartment and replace the battery. Close 

battery cover.

Note: Batteries should only be replaced in a safe area using the 

battery type specified in this instruction manual. Old batteries should 

be disposed in accordance with local regulations.

Errors that may occur: A message of 

“CAL”

 and 

“Err”

 means an 

internal error has occurred and Checktemp

®

 is not guaranteed to 

be within its specifications but may still be operational. If CalCheck

®

 

verifies Checktemp

®

 is still accurate and the display shows 

“---” 

the temperature being measured is out of the specified range or 

connection to the probe has been broken. If a message of 

“Err”

 occurs 

then a calibration data read error has occurred. A message of 

“Er1”

 

is caused by an unexpected device reset. A message of 

“Er2”

 occurs 

if internally a temperature measurement malfunctions. After these 

messages occur it is possible to be fixed by a reset. Turn Checktemp

®

 

OFF and then ON with the ON/OFF button to reset. If the error message 

continues to show, please contact your local Hanna Instruments Office.

Checktemp

®

 is guaranteed for a period of one year after date of 

purchase against defects in workmanship and materials when 
used for their intended purpose and maintained according to 
instructions. This warranty is limited to repair or replacement free 
of charge. Damage due to accidents, misuse, tampering or lack of 
prescribed maintenance is not covered. If service is required, 
contact your local Hanna Instruments Office. If under warranty, report 
the model number, date of purchase, serial number and the nature 
of problem. If the repair is not covered by the warranty, you will be 
notified of the charges incurred. If the instrument is to be returned to 
Hanna Instruments, first obtain a Returned Goods Authorization 
number from the Technical Service department and then send it with 
shipping costs prepaid. When shipping any instrument, make sure it is 
properly packed for complete protection.

Before using Hanna Instruments products, make sure that they are 

entirely suitable for your specific application and for the environment 

in which they are used. Any variation introduced by the user to the 

supplied equipment may degrade the meter's performance. To avoid 

damages or burns, do not put the meter in microwave oven. For yours 

and the meter's safety do not use or store the meter in hazardous 

environments.