94
Molybdenum
M O L Y B D E N U M
M O L Y B D E N U M
M O L Y B D E N U M
M O L Y B D E N U M
M O L Y B D E N U M
SPECIFICATIONS
Range
0.0 to 40.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.3 mg/L ±5% of reading
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 420 nm
Method
Adaptation of the mercaptoacetic acid method.
The reaction between molybdenum and the re-
agents causes a yellow tint in the sample.
REQUIRED REAGENT
Code
Description
Quantity
HI 93730
A
-0
Reagent A
1 packet
HI 93730
B
-0
Reagent B
1 packet
HI 93730
C
-0
Reagent C
1 packet
REAGENT SETS
HI 93730-01
Reagents for 100 tests
HI 93730-03
Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Molybdenum on the
secondary LCD by pressing PROGRAM
W
and
V
.
• Fill the cuvet with 10 mL of unreacted sample
(up to the mark) and replace the cap.
• Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
• Press ZERO and “SIP” will blink on the display.
10 mL