SPECIFICATIONS
Range
0.0 to 10.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.1 mg/L ±3% of reading
Typical EMC
±0.1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference
filter @ 525 nm
Method
Adaptation of the
EPA DPD method 330.5.
The
reaction between free chlorine and the DPD re-
agent causes a pink tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93701-0
DPD
1 packet
HI 93734
B
-0 Free & Total Chlorine HR Reagent
5 mL
HI 93734
C
-0 Total Chlorine HR Reagent
3 drops
REAGENT SETS
HI 93734-01
Reagents for 100 tests
HI 93734-03
Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Total Chlorine HR on
the secondary LCD by pressing PROGRAM
W
and
V
.
•
Add to the cuvet 5 mL of HI 93734B reagent
by means of the 5 mL syringe.