Hanna Instruments BL 7917 Скачать руководство пользователя страница 7

13

12

Probe connections

Connect the pH/ORP electrode to the BNC socket of the pump.

The foot valve assembly (

HI 721005

)

should always hang vertically and not
lay horizontal on the bottom of the tank
or drum.
A vertical assembly will ensure that the
valve is positioned properly and prevent
loss of prime.
For the US standard installations, use
PVC adapters to connect the suction and
discharge valves to the PVC pipe.

To ensure maximum performance, check the voltage at the point of
supply to verify that it is sufficient. It is recommended that you install a
1 Amp circuit breaker between the pump and the power supply. This
will give additional protection to the internal circuit and provide a
convenient way to disconnect the power supply prior to servicing the
pump, if needed.

Injection Point

• Choose an injection point that allows you to mount the

injection valve assembly vertically.

• The spring in the injection valve assembly (

HI 721004

) adds

approximately 1.5 bar of back pressure. If pumping into a
high back pressure, the spring should be removed.

Other Considerations

• If you are mounting the system to a wall, column, etc., be sure

it is strong enough to support the weight of the entire system.

• The ambient temperature of the pump, when in operation,

should be between 0 and 50°C (32 to 122°F) and should be
protected from direct exposure to outdoor elements (direct sun-
light, rain, extreme temperatures, high humidity, etc.).

• Generally speaking, the shorter the suction distance, the more

efficient the pump operates.

• The pump should be placed in a conventional location that will

allow easy access to the control and connections. It should be
placed so that regular visual inspections of the connections and
hoses are facilitated.

Vertical Surface Mounting

Once you have selected the best installation site, simply screw or bolt
the unit into a wall or mounting panel above the chemical feed tank.
The 4 mounting screw holes on the pump will accommodate up to a 5
mm (3/16") screw or bolt (remember to use heavy screws or bolts to
secure the system). Be sure you do not over tighten and cause excessive
stress on the mounting holes.
Make sure to leave a slight overhang in front to allow for the
connection cable.

Power Supply

Connect the power cord to the female socket of the pump and by doing
so also grounding it. The power socket contains a 1 A/250 V fuse.
Since there is no on/off switch, it is suggested to install an outside
switch.

U

NION

 C

ONNECTION

A

DAPTER

D

ISCHARGE

 V

ALVE

P

UMP

S

UCTION

 V

ALVE

A

DAPTER

F

OOT

 V

ALVE

F

ILTER

Diagram for Rigid Pipe Hose

-

Permanent Connection using 3/8" PVC
pipe

All piping for the pump feed and dis-
charge should be plumbed to the location
of the pump.
The threads on both valve assemblies
allow the use of standard 3/8" (Euro-
pean) pipe fittings for permanent pipe
connections.

Содержание BL 7917

Страница 1: ...Instruction Manual BL 7916 BL 7917 pH and ORP Measuring Dosing System w w w h a n n a i n s t c o m...

Страница 2: ...near chemical dosing pumps When pumping chemicals make sure all tubes are securely attached to the fittings It is recommended that tubing is shielded to prevent possible injury in case of rupture or a...

Страница 3: ...ch of children and pets All pumps undergo stringent tests to ensure that they comply with their stated specifications and are calibrated at the maximum rated pressure GENERAL DESCRIPTION The BL 7916 a...

Страница 4: ...P electrode 11 Overheating LED 12 Power Socket and Fuse Holder 13 Cable Glands FUNCTIONAL DESCRIPTION BL 7916 1 Discharge Valve Assembly 2 Pump head 3 Suction Valve Assembly 4 Hose 5 Liquid Crystal Di...

Страница 5: ...142 x 181 mm 8 7 x 5 6 x 7 1 Weight Approximately 5 kg 11 lb Range 0 00 to 14 00 pH Resolution 0 01 pH Accuracy 20 C 68 F 0 01 pH Typical EMC Deviation 0 1 pH InputImpedance 10 Ohm Dosage Proportional...

Страница 6: ...r permanent installation The pump can be mounted directly on a wall or tank see page 10 for the specific mounting dimensions Power Requirements BlackStone pumps are designed to operate to specificatio...

Страница 7: ...should be protected from direct exposure to outdoor elements direct sun light rain extreme temperatures high humidity etc Generally speaking the shorter the suction distance the more efficient the pu...

Страница 8: ...nd discharge valves located on the pump head should not be inter changed as they are different internally The discharge valve is fitted with a valve guide and will not function properly if used on the...

Страница 9: ...INSTALLATIONS Flooded Suction Installation Suggested Installation for consis tent output when using a low stroke rate Also suggested for highly viscous chemicals A slight suction pressure avoids self...

Страница 10: ...pump works at 100 of the rating when the difference between measure and setpoint is more than 150 mV BL 7917 or 1 5 pH BL 7916 It goes into proportional dosing for values less than the above You can...

Страница 11: ...ump is stopped when the chemicals are exhausted MANUAL An auxiliary switch can be connected to these terminals to force the unit to pump whatever value is measured by the controller This might be need...

Страница 12: ...nk This vertical distance should not exceed 1 5 meters 5 feet Either lower the pump or raise the feed tank Check the pump head suction and discharge valves for blockage Pump flow rate is reduced Check...

Страница 13: ...Remove the valves from the pump head the injection fitting and the feed Keep the suction and discharge valves separated as they are not interchangeable Disassemble each valve and clean it with a neutr...

Страница 14: ...constantly in a well filled with the sample stream or tank When not in use replace the solution in the protective cap with a few drops of HI 70300 storage solution Follow the Preparation Procedure ab...

Страница 15: ...he cleaning procedures rinse the electrode thoroughly with distilled water drain and refill the reference chamber with fresh electrolyte not necessary for gel filled electrodes and soak the electrode...

Страница 16: ...or reducing power of a solution and are commonly expressed in mV Oxidation may be defined as the process during which a molecule or an ion loses electrons and reduction as the process by which electr...

Страница 17: ...thane Tetraethyl Lead Tetralin Tin Salts Vegetable Oils Vinegar Water Acid Mine Water Fresh Water Distilled Water Salt Whiskey Wines Zinc Chloride Zinc Sulfate Adipic Acid Alcohol Amyl Alcohol Diaceto...

Страница 18: ...L Oxidizing pretreatment solution 500 mL MAINTENANCE SOLUTIONS HI 70300M Storage solution 230 mL botte HI 70300L Storage solution 500 mL bottle HI 7061M General cleaning solution 230 mL botte HI 7061L...

Страница 19: ...5M KCl AgCl electrolyte solution 4x50 mL for single junction electrodes HI 7072 1M KNO3 electrolyte solution 4x50 mL HI 7082 3 5M KCl electrolyte solution 4x50 mL for double junction electrodes RECOMM...

Страница 20: ...e sure that they are entirely suitable for the environ ment in which they are used Operation of these instruments in residential areas could cause unacceptable interfer ence to radio and TV equipment...

Страница 21: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Отзывы: