SETPOINT ADJUSTMENTS

• Move the display selector switch to SET.

• With a small screwdriver adjust the SET COARSE trimmer to

display a value close to the desired setpoint e.g. 7.00.

• Adjust the SET FINE trimmer to display the exact setpoint value.

• Once the desired setpoint is achieved, move the display selector

switch back to MEASURE.

ACID OR BASE SELECTION (BL 7916 only)

If an acid chemical is to be dosed, move the
ACID/BASE switch to ACID. This means that
the pump will dose when pH measurements
are higher than the setpoint.
If an alkaline chemical is to be dosed, move
the ACID/BASE switch to BASE. This means
that the pump will dose when pH measure-
ments are lower than the setpoint.

REDUCING OR OXIDIZING SELECTION (BL 7917 only)

If a reducing chemical is to be dosed, move
the RED/OXID switch to RED. This means that
the pump will dose when redox measure-
ments are higher than the setpoint.
If an oxidizing chemical is to be dosed, move
the RED/OXID switch to OXID. This means
that the pump will dose when the redox
measurements are lower than the setpoint.
The wires and cables needed for all connections can be fed through
the two cable fittings located below the terminals. It is important to
attach these cables properly because pump vibrations could loosen
them. Replace the transparent cover and fasten the screws.

For example, a level controller can be placed in the tank of the liquid
being dosed so that the pump is stopped when the chemicals are
exhausted.

MANUAL

An auxiliary switch can be connected to these terminals to force the
unit to pump whatever value is measured by the controller. This
might be needed for example when the pump needs to be primed.

AUXILIARY CONTACT

The controller can drive a relay when the measure overtakes the
setpoint. The contact available on the terminals can be normally open
or closed. It can be used as a switch to drive an external mixer when
the pump is dosing, or can activate an auxiliary pump or any other
device. With 

BL 7916

, the contact can be selected to remain open

when the pH value is within the set limits (the pump is not dosing)
and closed when the pump is dosing or viceversa.
By shorting the Common (middle) terminal with the NO or NC
terminal on the driving circuit inside the controller section, a normally
open or a normally closed state can be achieved, as shown below:

ALARM CONTACT

If the reading drifts from the setpoint by more than ±2.00 pH in

BL 7916

 and ±200 mV in 

BL 7917

, an external alarm can be

activated. The contact is rated at 2 amps at 220VAC. No power is
supplied through the system and it is an open/closed contact only.
With 

BL 7916

, the contact can be selected to be closed when the

alarm is activated and open when the alarm is off or viceversa. Short
the middle terminal located on the driving circuit inside the controller
section with the NO or NC terminal (see above).

4-20  mA ISOLATED OUTPUT

A 4-20 mA proportional output is available in these connections:

4 mA=0 pH; 20 mA=14 pH (BL 7916)

and

4 mA=-999 mV; 20 mA=+999 mV (BL 7917).

The output is optically isolated.

Содержание BL 7917

Страница 1: ...Instruction Manual BL 7916 BL 7917 pH and ORP Measuring Dosing System w w w h a n n a i n s t c o m...

Страница 2: ...near chemical dosing pumps When pumping chemicals make sure all tubes are securely attached to the fittings It is recommended that tubing is shielded to prevent possible injury in case of rupture or a...

Страница 3: ...ch of children and pets All pumps undergo stringent tests to ensure that they comply with their stated specifications and are calibrated at the maximum rated pressure GENERAL DESCRIPTION The BL 7916 a...

Страница 4: ...P electrode 11 Overheating LED 12 Power Socket and Fuse Holder 13 Cable Glands FUNCTIONAL DESCRIPTION BL 7916 1 Discharge Valve Assembly 2 Pump head 3 Suction Valve Assembly 4 Hose 5 Liquid Crystal Di...

Страница 5: ...142 x 181 mm 8 7 x 5 6 x 7 1 Weight Approximately 5 kg 11 lb Range 0 00 to 14 00 pH Resolution 0 01 pH Accuracy 20 C 68 F 0 01 pH Typical EMC Deviation 0 1 pH InputImpedance 10 Ohm Dosage Proportional...

Страница 6: ...r permanent installation The pump can be mounted directly on a wall or tank see page 10 for the specific mounting dimensions Power Requirements BlackStone pumps are designed to operate to specificatio...

Страница 7: ...should be protected from direct exposure to outdoor elements direct sun light rain extreme temperatures high humidity etc Generally speaking the shorter the suction distance the more efficient the pu...

Страница 8: ...nd discharge valves located on the pump head should not be inter changed as they are different internally The discharge valve is fitted with a valve guide and will not function properly if used on the...

Страница 9: ...INSTALLATIONS Flooded Suction Installation Suggested Installation for consis tent output when using a low stroke rate Also suggested for highly viscous chemicals A slight suction pressure avoids self...

Страница 10: ...pump works at 100 of the rating when the difference between measure and setpoint is more than 150 mV BL 7917 or 1 5 pH BL 7916 It goes into proportional dosing for values less than the above You can...

Страница 11: ...ump is stopped when the chemicals are exhausted MANUAL An auxiliary switch can be connected to these terminals to force the unit to pump whatever value is measured by the controller This might be need...

Страница 12: ...nk This vertical distance should not exceed 1 5 meters 5 feet Either lower the pump or raise the feed tank Check the pump head suction and discharge valves for blockage Pump flow rate is reduced Check...

Страница 13: ...Remove the valves from the pump head the injection fitting and the feed Keep the suction and discharge valves separated as they are not interchangeable Disassemble each valve and clean it with a neutr...

Страница 14: ...constantly in a well filled with the sample stream or tank When not in use replace the solution in the protective cap with a few drops of HI 70300 storage solution Follow the Preparation Procedure ab...

Страница 15: ...he cleaning procedures rinse the electrode thoroughly with distilled water drain and refill the reference chamber with fresh electrolyte not necessary for gel filled electrodes and soak the electrode...

Страница 16: ...or reducing power of a solution and are commonly expressed in mV Oxidation may be defined as the process during which a molecule or an ion loses electrons and reduction as the process by which electr...

Страница 17: ...thane Tetraethyl Lead Tetralin Tin Salts Vegetable Oils Vinegar Water Acid Mine Water Fresh Water Distilled Water Salt Whiskey Wines Zinc Chloride Zinc Sulfate Adipic Acid Alcohol Amyl Alcohol Diaceto...

Страница 18: ...L Oxidizing pretreatment solution 500 mL MAINTENANCE SOLUTIONS HI 70300M Storage solution 230 mL botte HI 70300L Storage solution 500 mL bottle HI 7061M General cleaning solution 230 mL botte HI 7061L...

Страница 19: ...5M KCl AgCl electrolyte solution 4x50 mL for single junction electrodes HI 7072 1M KNO3 electrolyte solution 4x50 mL HI 7082 3 5M KCl electrolyte solution 4x50 mL for double junction electrodes RECOMM...

Страница 20: ...e sure that they are entirely suitable for the environ ment in which they are used Operation of these instruments in residential areas could cause unacceptable interfer ence to radio and TV equipment...

Страница 21: ...2 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 01...

Отзывы: