Tips for an Accurate Measurement

• Ensure the sample does not contain any debris.
• Whenever the cuvette is placed into the checker, it must be 

dry outside and free of fingerprints, oil and dirt. 

• Wipe the cuvette thoroughly with 

HI731318

 microfiber 

cleaning cloth or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles, causing higher 

readings. To obtain accurate measurements, remove such 

bubbles by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent 

has been added, as accuracy will be affected.

• Discard the sample immediately after the reading has been 

taken or the glass might become permanently stained.

Battery Replacement

To save the battery, the checker shuts down after 10 minutes of non-use and 

2 minutes after reading.

A fresh battery lasts for a minimum of 5000 measurements. When the battery is 

drained, the instrument displays “bAd” then ”bAt”, and turns off. 

To replace the battery, follow the next steps:

1. Press and hold the ON/OFF button to turn the checker off.

2. Turn the instrument upside down and use a screwdriver to unfasten the screw and 

remove the battery cover.

3. Remove the old battery, replace it with a new 1.5V AAA battery, inserting the 

negative end first.

4. Replace the battery cover, fasten and tighten the screw.

www.

.com

[email protected]

1.800.561.8187

Содержание Alkalinity HI775

Страница 1: ...HI775 Alkalinity INSTRUCTION MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...rvice is required contact your local Hanna Instruments Office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the...

Страница 3: ...d the sample immediately after the reading has been taken or the glass might become permanently stained Battery Replacement To save the battery the checker shuts down after 10 minutes of non use and 2...

Страница 4: ...3755 53 Chlorine remover reagent Certification All Hanna Instruments conform to the CE European Directives Disposal of Electrical Electronic Equipment The product should not be treated as household wa...

Страница 5: ...Alkaline battery 1 pc Instruction manual Note Save all packing material until you are sure that the Checker HC handheld colorimeter works correctly Any damaged or defective item must be returned in i...

Страница 6: ...nish blue to develop This checker has been developed to work with fresh water samples Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Battery type 1 5V AAA Alkaline Auto shut off After 10 m...

Страница 7: ...chlorine interference add one drop of HI93755 53 Chlorine Remover to the unreacted sample Remove the cuvette and unscrew the cap Using a 1 mL syringe carefully add exactly 1 mL of HI775S Alkalinity re...

Страница 8: ...inverted Swap the cuvettes and repeat the measurement UnderRange Ablinking 0 indicatesthatthesampleabsorbs less light than the zero reference Check the measurement procedure and make sure to use the...

Отзывы: