background image

Errors & Warnings

The checker shows clear warning messages when erroneous conditions appear and 
when measured values are outside the expected range. The information below 
provides an explanation of the errors and warnings, and the recommended action 
to be taken.

Light High:

 There is an excess amount of ambient light 

reaching the detector. Please check the preparation of the zero 
cuvette.

Light Low:

 There is not enough light to perform a measurement. 

Please check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvettes are 

inverted. Swap the cuvettes and repeat the measurement.

Under Range:

 A blinking “0”

 

indicates that the sample absorbs 

less light than the zero reference. Check the measurement 
procedure and make sure to use the same cuvette for reference 
(zero) and measurement.

Over Range:

 Maximum concentration value displayed blinking 

indicates the measured value is outside the limits of the method.
Verify that the sample does not contain any debris. Dilute the 
sample and repeat the measurement.

Battery Low:

 Battery level is too low for the checker to function 

properly. Replace the battery with a new one.

Drained Battery:

 The battery is drained and must be replaced. 

Replace the battery with a new one and restart the checker.

www.

.com

[email protected]

1.800.561.8187

Содержание Alkalinity HI775

Страница 1: ...HI775 Alkalinity INSTRUCTION MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...rvice is required contact your local Hanna Instruments Office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the...

Страница 3: ...d the sample immediately after the reading has been taken or the glass might become permanently stained Battery Replacement To save the battery the checker shuts down after 10 minutes of non use and 2...

Страница 4: ...3755 53 Chlorine remover reagent Certification All Hanna Instruments conform to the CE European Directives Disposal of Electrical Electronic Equipment The product should not be treated as household wa...

Страница 5: ...Alkaline battery 1 pc Instruction manual Note Save all packing material until you are sure that the Checker HC handheld colorimeter works correctly Any damaged or defective item must be returned in i...

Страница 6: ...nish blue to develop This checker has been developed to work with fresh water samples Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Battery type 1 5V AAA Alkaline Auto shut off After 10 m...

Страница 7: ...chlorine interference add one drop of HI93755 53 Chlorine Remover to the unreacted sample Remove the cuvette and unscrew the cap Using a 1 mL syringe carefully add exactly 1 mL of HI775S Alkalinity re...

Страница 8: ...inverted Swap the cuvettes and repeat the measurement UnderRange Ablinking 0 indicatesthatthesampleabsorbs less light than the zero reference Check the measurement procedure and make sure to use the...

Отзывы: