3. Installation
3.1 Packing Inspection
3.2 Installation
3.2.1 Selecting the Location
▶ You can get contact details on the packing box or manual.
• Check packing conditions carefully, before unpacking.
• Contact Hanil Scientific Inc. immediately if damages found.
• Check the delivery for completeness.
Installation on hard and flat ground.
• Centrifuge should be installed on hard and flat place.
• If the centrifuge is installed in an inclined place, the shaft may be bent
due to the weight of the rotor.
Good ventailation.
• For air circulation and safety, maintain a 30cm“clearance envelope”
around the centrifuge while the rotor is spinning.
Constant temperature/humidity
• Centrifuge equipped with the sensitive electronic software which is
fragile with humidity and temperature.
• Must avoid direct ray or heater and be put in the ambience of controlled
temp.and air.
Avoid the corrosive gas
• Install the centrifuge in a place where corrosive gas is not generated.
• Sulfur dioxide gas and chlorine gas may cause corrosion.
Leveling
• The device should be operated on a flat surface.
Use a leveler tool to check the device is level.