![Hanil Component R12 Скачать руководство пользователя страница 19](http://html2.mh-extra.com/html/hanil/component-r12/component-r12_operation-manual_3605064019.webp)
Technical Support :+82-2-3452-8966 / [email protected]
19/23
5.3 Care Instructions for Refrigerated device
5. Maintenance
[Open the lid after centrifugation]
1. Wipe up the moisture in the chamber with a soft cloth.
2. Leave the centrifuge lid open after centrifugation.
[Remove the condensation water]
1. There is a drain hose at the bottom of the device.
Regularly check and remove the condensation water from the drain hose.
[Remove dust]
1. Remove any dust from the ventilation slits regularly.
Drain Hose
ventilation Slots