3.

 

When ready for treatment that will create airborne aerosol, adjust the rigid flexible hose and 
hood directly in front of the patient’s mouth approximately 3” to 4” from the oral cavity. 

4.

 

The green lighted rocker switch should always remain in the ON position and the variable speed 
knob adjusted to the highest position for maximum suction.  The variable speed knob can be 
adjusted to lower suction as per user preference. 

5.

 

Use  the  “hands  free”  foot  control  on/off  switch  during  procedures  to  reduce  the  risk  of 
contamination to gloved hands. 

6.

 

As you are performing work on a patient, you may adjust the proximity of the hood opening for 
maximum efficiency. 

a.

 

Note, keep the rigid hose as compressed as possible to avoid slumping or hose movement 
during procedures. 

7.

 

NOTE: Never pull the 42ESU by the hose as it will cause the hose, always use the black handle 
located on top of the unit. 

 

Recommended Cleaning Instructions

 (Does not supersede ADA, CDC or OSAP disinfection 

guidelines): 

After each use

, the hood opening, and rigid flexible hose should be cleaned. 

1.

 

Make sure the unit is turned off and the hose is fully compressed and in a standing upright 
position. 

2.

 

Using approved disinfection wipes, wipe down the inside and outside of the hood opening 
thoroughly.  Using a second wipe, wipe down the outside of the hose thoroughly.   

3.

 

Turn the unit on with variable speed knob turned to the lowest setting to reduce CFM.  Spray 
the inside of the hood opening for 5 seconds with an aerosol disinfection agent that is greater 
than 70% alcohol.  The suction will draw down the disinfecting spray into the hood opening to 
cover the inside of the hood opening and rigid flexible hose from top to bottom.  Shut the unit 
off and reset the variable speed knob to its appropriate setting. 

4.

 

Any additional required cleaning in accordance with CDC, ADA or OSAP guidelines.

 

 

Filter Maintenance: 

1.

 

Make sure the unit is turned off and unplugged when performing filter maintenance. 

2.

 

The 

HANDLER 42ESU

 has two HEPA filters. 

 

 

A disposable H11 HEPA filter bag is located in the front of the unit.   

 

a.

 

This filter should be changed on a regular basis.  Suggested filter replacement is once 
monthly, possibly sooner based upon heavy use.  The part number to order is 62-IIDF, set 
of 3 filters. 

b.

 

ALWAYS WEAR APPROPRIATE PPE WHEN ACCESSING AND REPLACING FILTERS. 

c.

 

To access this disposable filter bag filter, remove the two thumbscrews in the front and 
remove the door. 

d.

 

Slowly  and  carefully  remove  the  used  filter  bag  and  dispose  of  it  according  to  CDC 
guidelines as it may contain contaminants. 

e.

 

Spray the inside front cavity generously with an aerosol disinfection agent that is greater 
than 70% alcohol and allow to dry. 

f.

 

Install a new bag making sure the rubber gasket is completely sealed to the suction inlet 
tube. 

g.

 

Put the front door back on and wipe down the unit. 

 

Содержание AERO Series

Страница 1: ...ools NEVER use this product to pick up flammable liquids or any liquids and do not use near flammable or combustible liquids as fire or explosion can occur ALWAYS operate the 42ESU with filters properly in place and inspected for any damages Debris can cause inhalation problems ALWAYS disconnect the 42ESU from a power source when removing or replacing filter bags or doing any sort of maintenance ...

Страница 2: ... to hold in place Plug the power cord into the foot switch and plug the foot switch into a wall outlet 5 Attach the rigid hose with factory hood to the suction inlet of the base unit using the hose assembly instruction file attached to the hose 6 Your HANDLER 42ESU is now ready for use 7 NOTE Never pull the 42ESU by the hose as it will cause the hose to lose rigidity Operating Instructions 1 The H...

Страница 3: ...horoughly 3 Turn the unit on with variable speed knob turned to the lowest setting to reduce CFM Spray the inside of the hood opening for 5 seconds with an aerosol disinfection agent that is greater than 70 alcohol The suction will draw down the disinfecting spray into the hood opening to cover the inside of the hood opening and rigid flexible hose from top to bottom Shut the unit off and reset th...

Страница 4: ... One Year Limited Warranty Handler Mfg LLC 612 North Avenue East Westfield NJ 07090 908 233 7796 Ph 908 233 7340 Fax info handlermfg com LIMITED WARRANTY Warranty period is 1 year for the 42ESU Extraoral Suction Unit Handler Mfg LLC warrants to the original purchaser only that its products will be free from defects in material or workmanship for the applicable period of time indicated above follow...

Отзывы: