Information panel
On the control panel at the bottom of the lift unit there are several symbols. The symbols lights up
when a function is activated. More information on each function is given below, and under chapter
Trouble-shooting.
Battery status
Battery status is indicated for 2 seconds when a lift with end point charging is activated, and for 15
minutes if the in-rail charging option is used.
Service/maintenance
Information concerning service and maintenance. This symbol is activated once the lift unit has
been used for 12 months.
Overload protection
RiseAtlas is equipped with overload protection, which prevents the lift from lifting more than the
maximum rated load if this happen. A symbol for overload and a warning triangle light up and a
signal is heard. If this symbol is activated, the patient is too heavy.
Read manual
Often, the book symbol will light up at the same time as another symbol lights up, indicating that
the user should read the manual before using the lift again.
Night light
RiseAtlas is equipped with night lighting, which will be activated by pressing and holding the ON
button on the hand control for 3 seconds. The LED lamp lights for 15 minutes if it is not inactivated
via the OFF button on the hand control.
Emergency stop function
This light is only operational with kit 70200047, which is an auxiliary 9V battery back up, The light
indicates when the 9v battery is too low and the battery should be changed. The light will reset
when the new battery is installed. This symbol will also light up each time the lift unit is switched on
via the hand control.
Warning
The warning triangle is activated, prompting the user to read the manual. Often, the
warning triangle will light up at the same time as another symbol lights up, indicating that the user
should read the manual before using the lift again. Normally, a technician must be
contacted when this symbol is activated.
Preventative service/maintenance
The symbol for service/maintenance lights up automatically when the lift unit has been used for 12 months.
Contact an authorized Handicare service technician for maintenance and resetting.