background image

27

S E R V I C E   M A N U A L

System

RoMedic

TM

1.15 Actuator Motor Wiring Trip

This occurs when the control system detects a fault in the wiring to either actuator motor. Check the motor connectors 

and wiring. If the trip is still present after the above checks have been made, then the Power Module may be defective.

1.16 Current Limit Active

This occurs when the control system operates above the Current Limit Threshold for a period of time greater than the 

Current Limit Time. It has been designed to notify the Healthcare Technician that the control system has operated  

outside of its programmed range.

1.17 High Temperature

This occurs when the control system reaches its Temperature Threshold and thus becomes too hot. The controller will 

not allow drive until the controller has cooled down. An entry is made in the system log each time the controller gets too 

hot and goes out of drive.

Содержание SystemRoMedic EvaDrive

Страница 1: ...d on the lift Always read the user manuals for all assistive devices used during a transfer Keep the user manual where it is accessible to users of the product Always make sure that you have the right...

Страница 2: ...ting 5 Exploded view 6 8 Spare parts 10 14 Changing spare parts 12 14 Adjusting Toe in Toe out 15 Periodic inspection protocol 16 Description of periodic inspektion 17 Periodic inspection instruction...

Страница 3: ...width adjustment Activate the system press the button for base width adjustment open the undercarriage to maximum width and then press the other button to reduce the width If the lift is functioning...

Страница 4: ...rs or alterations are made by authorized personnel Never attempt to lift the mobile lift unit by the handlebar Ensure that there are no obstacles or people in the way of the lift Never stand on the li...

Страница 5: ...electrical system Restart EvaDrive If there is still a problem contact Handicare Red light battery scale blink Restart Contact Handicare If unusual sounds are heard Try to locate the source of the sou...

Страница 6: ...IN 985 CL6 M10 FZB 1 11 90000601 01 Washer DIN125 10 5x20x2 fzb 2 12 90000660 01 Nut cover MSM10 2 13 90000789 00B Screw MLC6S DIN7984 8 8 M5x8 2 4 5 C D 4 6 7 8 A B This drawing and any information o...

Страница 7: ...Internordic 4 14 90004722 01 Eva drive sleeve 4 15 90004721 01 Eva drive tension bracket 2 16 90005031 01A Eva drive support welded 1 17 90001674 01 Screw MC6S DIN912 8 8 M8x16 fzb 4 18 90004471 01 S...

Страница 8: ...0 Cover 20mm grey 1 11 90003072 01 Handle Bracket Eva Drive 1 12 90001674 01 Screw MC6S DIN912 8 8 M8x16 fzb 2 C D 7 8 A B This drawing and any information or descriptive matter set out hereon are the...

Страница 9: ...straight mini fit 1 2 90005089 01 EvaDrive actuator cable short angled micro fit 1 3 90005091 01 EvaDrive Emergency Stop Cable 1 4 90005092 01 EvaDrive battery cable 1 5 90003269 01 Eva Drive Handle 1...

Страница 10: ...90004212 01 Actuator LA30 2 way 478mm 301358 00 1 2 90005060 01 Nut ML6M DIN439 ISO4035 M10 fzb 1 3 90004723 03 Eva drive actuator stearing pin 1 D E F C 1 2 3 4 B A 3 2 1 5 C D 4 6 7 8 A B This drawi...

Страница 11: ...80100187 Cable actuator LA23 EvaDrive 1 pce 1 x Cable actuator LA23 80100189 Cable Emergency stop EvaDrive 1 pce 1 x Cable Emergency stop 80100190 Charger Li Ion EvaDrive 1 pce 1 x Charger Li Ion Eva...

Страница 12: ...y slide battery back into place Reattach battery cable and switch ON Readjust handle position 80100174 Cover leg EvaDrive Replace cover leg by pressing it into back of leg profile use LocTite715 glue...

Страница 13: ...l key in the keyway Fit the spacer and tighten a new lock nut firmly Rout the cable and reattach the connector to the HUB motor Secure HUB motor cable with cable ties Note cable tension Reattach cover...

Страница 14: ...itch ON 80100189 Cable Emergency stop EvaDrive Set battery switch OFF Disconnect battery Detach cover The cover is secured by pads of velcro Detach emergency stop cable connector from control box VR2...

Страница 15: ...am but do not remove them 3 Loosen all four screws connecting the center plate with the brackets but do not remove them 4 Straighten one of the wheels by hand Note that the loose bracket moves relativ...

Страница 16: ...amage Inside the unit Cables Weare and damage Noise Accessories Documentation Instructions User manuals Lift type Contract no s n Name Version Address Prod year Sling bar id Situation of use Home Inst...

Страница 17: ...17 S E R V I C E M A N U A L SystemRoMedic TM A B D C E F G Description for periodic inspection...

Страница 18: ...and then try to drive backwards and forwards Nothing should happen then press the up and down button If nothing happens the emergency stop buppen functions properly Inspect the handle Check screws nu...

Страница 19: ...irt and hair Do not use cleaning agents containing phenol or chlorine as this could damage the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should...

Страница 20: ...ss than 10 though not more than 2 min Nav motor 24V DC Mono Shaft 180 W Material Aluminium steel Emergency lowering Manual Castors 8 inch 203 mm IP class IP X4 Weight Eva450Drive 127 87 lbs 58 kg Eva6...

Страница 21: ...o the degree of protection against electric shock The device is intended for indoor use Class lI equipment Conforms to ISO 10535 2006 with ANSI AAMI ES 60601 1 2005 and is certified to ISO 10535 2006...

Страница 22: ...50Drive 25 0 9 427 16 8 1307 51 4 1450 57 1 240 9 4 900 1100 35 4 43 3 All measurement are in mm inch Tolerance 5 mm 0 2 When the handle is placed 1000 mm from floor Dimensions 22 S E R V I C E M A N...

Страница 23: ...M A N U A L SystemRoMedic TM Eva600Drive 806 31 7 395 15 5 25 0 9 432 17 1408 55 4 1566 61 6 240 9 4 900 1100 35 4 43 3 All measurement are in mm inch Tolerance 5 mm 0 2 When the handle is placed 100...

Страница 24: ...7 Dual Joystick Displaced Refer to section 1 11 4 N 7 or 8 Other Possible Control System Trip 1 7 1 8 3 Y 8 or 8 Other Possible Control System Trip 1 7 1 8 1 1 Trip Type 1 Low Battery Voltage This occ...

Страница 25: ...eck the right motor connectors and wiring If the trip is still present after the above checks have been made then the Power Module may be defective The VR2 control system may be programmed to exchange...

Страница 26: ...before the blink finishes If the problem persists refer to section 1 7 1 12 Communications Error The most likely cause of a communications error is a defective cable between the Power Module and the...

Страница 27: ...occurs when the control system operates above the Current Limit Threshold for a period of time greater than the Current Limit Time It has been designed to notify the Healthcare Technician that the co...

Страница 28: ...oMedic offers improvement of both work environment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the manual The most...