15
S E R V I C E M A N U A L
System
RoMedic
TM
Adjusting Toe-in/Toe-out
1. Remove the cover by gently pulling it
at the four Velcro points
2. Loosen all four screws securing ONE
of the brackets to the center beam, but
do not remove them.
3. Loosen all four screws connecting the
center plate with the brackets, but do
not remove them
4. Straighten one of the wheels by hand.
Note that the loose bracket moves
relative to the center beam.
5. With the wheel straight, tighten the two center plate screws.
6. Straighten the other wheel by hand.
7. Tighten the two remaining center plate screws.
8. Tighten all four screws loosened in (2).