Emergency lowering
-Verify the function of emergency lowering.
Charging
-Verify that charging is operating. Yellow lamp on the controllbox should light up when charging.
Product label
- Check the serial number so that the label is there and readable. Note the serial number on the protocol.
Test loading
Test loading of MiniLift160 Classic is done with 120 kg weight, which corresponds to a patient weight of 160 kg.
Test loading with loose weights
NOTE! Both hooks must bear the weight. When one weight is used, the weight
must be distributed over both hooks. Ensure that the weight pack is well clear of
the actuator. Weights must be attached to the hooks as shown in picture.
Remove the lower-leg support and footrest before test loading; this allows
more space for the weight that is to be lifted.
NOTE! Replace the lower-leg support and footrest after test loading.
Move the boom to the outermost position and apply 120 kg.
- Run the actuator 160 mm up/down as shown in picture. The boom is loaded
with 120 kg and placed in the outermost position.
Listen for unusual sounds during raising/lowering.
Connector hooks
- Lift up to 120 kg clear of the floor and wait for 20 sec.
NOTE! Load the hooks according to picture.
The base is loaded with 160 kg.
- Place 160 kg on the footrest, within the frame for foot placement. Run the base-width
adjustment in/out and listen for unusual sounds.
Mechanical emergency lowering
-Verify the function of emergency lowering.
Documentation
- Check that the manual is available and up-to-date.
Accessories
-Check the accessories to the lift.
To reset the service counters press both lift buttons on the hand control at the same time for 5 seconds.
An audio signal will indicate that the timer has been reset.