EN-9
Calibrating the QC890 Verifier
Calibration is a requirement of ISO/IEC standards. Refer to the QC890 Verifier
User’s Guide or help file for additional calibration information. When calibrating
the QC890 Verifier system, the QC890 Verifier must be connected to the Quick
Check ImageData software. During calibration, it is important to align the center
of the barcode symbol to the center arrows and keep the unit parallel to the
symbol. See directions on page EN-8.
1. Place the verifier on top of the reflectance page that is included in your kit.
Use the symbol alignment window to center the reflectance page.
2. Press the scan button on top of the unit to turn on illumination to aid in
alignment.
3. Press the scan button on top of the unit a second time to read the barcode
symbol. The QC890 Verifier automatically reads the symbol and performs
calibration. If the calibration process is successful, the following screen
displays.
4. Enter your name and the location where the calibration was performed and
click on
OK
. The following screen displays.
5. If the unit does not calibrate, contact Hand Held Products Technical Support.