background image

29

Subject to change without notice

positive going edges must be chosen. In the case of sync pulses
below the field/line, the leading edge is negative and consequently
the slope selection must be set for falling edges. Since the invert
function may cause a misleading display, it must not be activated.

On the 2 ms/div setting and field

 TV 

triggering selected, 1 field is

visible if a 50 fields/s signal is applied. If the holdoff control is in
fully ccw position, it triggers without line interlacing affects
caused by the consecutive field.

The display can be expanded by switching on the

 X-MAG. x10

function so that individual lines are recognizable. Commencing
with a frame synchronizing pulse, the display can also be expanded
with the knob

 TIME/DIV. 

But note that this  can result in an

apparently unsynchronized display as each frame (half picture)
triggers. This is due to the off set of half a line between frames.

The influence of the integrating network which forms a trigger
pulse from the vertical sync pulses may become visible under
certain conditions. Due to the integrating network time constant
not all vertical sync pulses starting the trace are visible.

On the 10µs/div setting and line TV triggering selected, approx.
1½  lines are visible. Those lines originate from the odd and even
fields at random.

The sync-separator-circuit also operates with external triggering.
It is important that the voltage range (0.3 V

pp

 to 3 V

pp

) for external

triggering should be noted. Again the correct slope setting is
critical, because the external trigger signal may not have the
same polarity or pulse edge as the test signal displayed on the
CRT. This can be checked, if the external trigger voltage itself is
displayed first (with internal triggering).

In most cases, the composite video signal has a high DC content.
With constant video information (e.g. test pattern or colour bar
generator), the DC content can be suppressed easily by AC input
coupling of the oscilloscope amplifier. With a changing picture
content (e.g. normal program), DC input coupling is recommended,
because the display varies its vertical position on screen with AC
input coupling at each change of the picture content. The DC
content can be compensated using the

 Y-POS. 

control so that the

signal display lies in the graticule area. Then the composite video
signal should not exceed a vertical height of 6div.

Line/Mains triggering (~)

The instrument specific information regarding this mode is part of
the section ”Controls and Readout” paragraph

 TRIG. MODE

[20]

.

This trigger mode is present if the READOUT indicates the

 “~”

symbol instead of the “trigger source”, “slope” and “coupling”
information. The trigger point symbol is inactive in line/mains
trigger mode as there is no direct amplitude relationship between
the trigger voltage and the signal voltage.

A voltage originating from mains/line (50 to 60 Hz) is used for
triggering purposes if the trigger coupling is set to 

~

. This trigger

mode is independent of amplitude and frequency of the Y signal
and is recommended for all mains/line synchronous signals. This
also applies  within certain limits,  to whole number multiples or
fractions of the line frequency. Line triggering can also be useful
to display signals below the trigger threshold (less than 0.5 div).
It is therefore particularly suitable for measuring small ripple
voltages of mains/line rectifiers or stray magnetic field in a circuit.

In this trigger mode the slope direction pushbutton selects the
positive or negative portion of the line/mains sine wave. The
trigger level control can be used for trigger point adjustment.

Magnetic leakage (e.g. from a power transformer) can be
investigated for direction and amplitude using a search or pick up
coil. The coil should be wound on a small former with a maximum
of turns of a thin lacquered wire and connected to a BNC
connector (for scope input) via a shielded cable. Between cable
and BNC centre conductor a resistor of at least 100 Ohm should
be series connected (RF decoupling). Often it is advisable to
statically shield the surface of the coil. However, no shorted turns
are permissible. Maximum, minimum, and direction to the
magnetic source are detectable at the measuring point by turning
and shifting the coil.

Alternate triggering

This trigger mode can be selected in

 DUAL 

mode by the aid of the

TRIG. SOURCE

 

[17] 

pushbutton

  (

please note ”Controls and

Readout”)

. In the case of chopped

 DUAL 

mode, selecting

alternate trigger mode automatically sets the instrument to
alternate

 DUAL 

mode.

Under

 TVL

,

 TVF 

and

 line/mains 

triggering conditions alternate

triggering can not be chosen. Thus only the following trigger
coupling modes are available in alternate trigger mode: 

AC

,

 DC

,

HF 

and

 LF

. The trigger point symbol is not displayed in alternate

trigger mode.

With alternate triggering it is possible to trigger two signals from
different frequency sources (asynchronous). In this case the
oscilloscope must be operated in

 DUAL 

alternate mode with

internal triggering and each input signal must be of  sufficient
height to enable trigger. To avoid trigger problems due to diffe-
rent

 DC 

voltage components,

 AC 

input coupling for both channels

is recommended.

The internal trigger source is switched in alternate trigger mode
in the same way as the channel switching system in 

DUAL

alternate mode, i.e. after each time base sweep. Phase difference
measurement is not possible in this trigger mode as the trigger
level and slope setting are equal for both signals. Even with 180°
phase difference between both signals, they appear with the
same slope direction.

If signals are applied with a high frequency ratio (difference), the
trace intensity then becomes reduced if the time base is set to
smaller time coefficients (faster sweep). This happens as the
number of sweeps does not increase because it depends on the
lower frequency signal, but with a faster sweep the phosphor
becomes less activated.

External triggering

The external trigger input is activated with the aid of the

 TRIG.

SOURCE [17] 

pushbutton (see ”Controls and Readout”), if the

trigger coupling is not set to line/mains trigger coupling. Then the
internal trigger source is deactivated. As the external trigger
signal applied at the

 TRIG. EXT 

socket normally has no relation

to the signal height of the displayed signal, the trigger point
symbol is switched off. The external trigger voltage must have a
minimum amplitude of 0.3 V

pp

 and should not increase above

3 V

pp

. The input impedance of the

 TRIG. EXT.

 socket is approx.

1 MOhm II 20 pF.
The maximum input voltage of the input circuit is 100V (DC+peak
AC). The external trigger voltage may have a completely different

Triggering and time base

Содержание HM504-2

Страница 1: ...furbished and purchasing new Financing options such as Financial Rental and Leasing are also available on application TMG will assist if you are unsure whether this model will suit your requirements C...

Страница 2: ......

Страница 3: ...O s c i l l o s c o p e H M 5 0 4 2 Manual English...

Страница 4: ...25 Phase comparison with Lissajous figures 26 Phase difference measurement in DUAL mode Yt 26 Phase difference measurement in DUAL mode 26 Measurement of an amplitude modulation 26 Triggering and tim...

Страница 5: ...trol cables and or radiation The device under test as well as the oscilloscope may be effected by such fields Although the in terior of the oscilloscope is screened by the cabinet direct radiation can...

Страница 6: ...signal level Time Base delay provide high X Magnification of any portion of the signal 100 MHz 4 Digit Frequency Counter Cursor and Automatic Measurement Save Recall Memories for Instrument Settings R...

Страница 7: ...ontrols Autoset automatic signal related parameter settings Save and Recall 9 instrument parameter settings Readout display of menu parameters cursors and results Autom Measurement Freq Period Vdc Vpp...

Страница 8: ...the handle can be set into diffe rent positions A and B carrying C horizontal operating D and E operating at different angles F handle removal T shipping handle unlocked STOP Attention When changing t...

Страница 9: ...nrequires the horizontal or inclined position STOP Do not obstruct the ventilation holes Specifications are valid after a 20 minute warm up period between 15 and 30 degr C Specifications without toler...

Страница 10: ...pprox 10 and the real voltage value is 11 higher The gain reduction error can not be defined exactly as the 3 dB bandwidth of the Y amplifiers differs between 50 MHz and 55 MHz When examining square o...

Страница 11: ...pe However if for example only the residual ripple of a high voltage is to be displayed on the oscilloscope a normal x10 probe is sufficient In this case an appropriate high voltage capacitor approx 2...

Страница 12: ...nt lowerinputsensitivity must be chosen until the vertical signal height is only 3 8 div With a signal amplitude greater than 160 Vpp and the deflection coefficient VOLTS DIV in calibrated condition a...

Страница 13: ...eristic resistance of the cable as a rule 50 Ohm Especially when transmitting square and pulse signals a resistor equal to the characteristic impedance of the cable must also be connected across the c...

Страница 14: ...d by reducing the READOUT intensity RO The INTENS FOCUS control knob adjusts the READOUT intensity Turning this knob clockwise increases and counter clockwisedecreasestheintensity Onlytheminimumrequir...

Страница 15: ...Rot with Int trace rotation by use of INTENS FOCUS knob and allows you to compensate the influence of the Earth s magneticfieldonthetracedeflection Toavoidmisadjustment due to unavoidable deflection...

Страница 16: ...selected This allows for a different slope setting for the triggered DELAY DTR time base mode 10 TR Trigger indicator LED The TR LED is lit in Yt mode if the triggering conditions are met for the firs...

Страница 17: ...the readout e g Y1 deflection coefficient input coupling The symbolizes calibrated measuring conditions and is replaced by the symbol in uncalibrated conditions VAR The vernier variable function is de...

Страница 18: ...e internal trigger signal originating from channel I or channel II or the external trigger signal can be chosen DUAL and Addition mode Briefly pressing opens the trigger source Pulldown menu with the...

Страница 19: ...s the way the trigger signal is connected to the trigger unit AC DC content suppressed DC peak value detection inactive HF high pass filter cuts off frequencies below approx 50 kHz trigger point symbo...

Страница 20: ...e trace start position which is present after switching over from sea to del This enables lower time deflection coefficient settings for signal expansion del In DELAY mode a trigger event does not sta...

Страница 21: ...PUT CH II BNC socket This BNC socket is the signal input for channel II The outer ground connection is galvanically connected to the instrument ground and consequently to the safety earth contact of t...

Страница 22: ...ifferent frequency responses of the trigger amplifier and Y amplifier cause a reduction of the measurement accuracy If relatively low frequency signals 20 Hz are present the measurement value continuo...

Страница 23: ...io X measurement causes the display of two long and one short CURSOR lines and is enabled inYt time base mode only The unit to be displayed must be selected by briefly pressing the UNIT 35 pushbutton...

Страница 24: ...es are displayed Briefly pressing SOURCE selects the channel and it s deflection coefficient for the measurement The CURSOR lines must be set to the signal according to the selected channel b DUAL mod...

Страница 25: ...ubmenus Main Menue 50 Ohm For high impedance loads Oscilloscope approx 1 M Ohm Digital Voltmeter approx 10 MOhm the output voltage is either 0 2 Volt DC or 0 2 Vpp AC square wave Under First Time Oper...

Страница 26: ...n generator provides a square wave signal with selectable frequencies and a very fast rise time 4 ns from the output socket below the CRT screen As the square wave signals are used for probe compensat...

Страница 27: ...the calibrator output via a suitable probe The voltage provided by the probe to a high impedance input 1 MOhm II 15 30 pF will correspond to the division ratio of the probe used 10 1 20 mVpp output S...

Страница 28: ...the phosphor at too high a brightness setting INTENS setting which causes either a lasting loss of brightness or in the extreme case complete destruction of the phosphor at this point Phase difference...

Страница 29: ...a sine signal Except when external trigger is used the trigger threshold can be stated as vertical display height in div at which the time base generator starts the display is stable and the trigger...

Страница 30: ...formation regarding this item can be noted in the Data Sheet The coupling setting TRIG MODE 20 and indication are described under Controls and Readout As the automatic triggering does not work below 2...

Страница 31: ...er threshold less than 0 5 div It is therefore particularly suitable for measuring small ripple voltages of mains line rectifiers or stray magnetic field in a circuit In this trigger mode the slope di...

Страница 32: ...gnals where the pulses alternately show a small difference of the peak amplitudes Only a very exact trigger level adjustment makes a single display possible The use of the holdoff control simplifies t...

Страница 33: ...osen in del and dTr mode is stored and automatically set after activating one of those modes If the stored time coefficient in del dTr mode was higher than the actual value in sea search mode the time...

Страница 34: ...me base mode I X x10 magnifier switched off I Optimum X and Y position settings I Trace and readout visible If DC trigger coupling had been selected AC trigger coupling will not be chosen and the auto...

Страница 35: ...w typical patterns displayed by the various components under test I Open circuit is indicated by a straight horizontal line I Short circuit is shown by a straight vertical line Testing Resistors If th...

Страница 36: ...or ground for the same terminal is then absolutely necessary A connection inversion effects a rotation of the test pattern by 180 degrees about the centre point of the scope graticule Pay attention t...

Страница 37: ...ation The oscilloscope is supplied with a serial interface for control purposes The interface connector 9 pole D SUB female is located on the rear of the instrument Via this bidirectional port the ins...

Страница 38: ...RECALL INPUT CHI X AC DC GND max 400Vp 1M II 18pF INPUT CHI X max 400Vp 1M II 18pF TRIG EXT max 100Vp INP Z x1 x10 AC DC GND x1 x10 50 MHz ANALOG OSCILLOSCOPE HM504 2 Instruments CAT I TRACE ROT CT GL...

Страница 39: ...37 Subject to change without notice Front Panel HM504 2...

Страница 40: ......

Страница 41: ...LEER...

Страница 42: ...chwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes Spectrum Analyzer P...

Отзывы: