Hameg HM303-6 Скачать руководство пользователя страница 8

8

Subject to change without notice

Type of signal voltage

With the 

HM303-6

, most repetitive signals in the frequency

range up to 

at least 35MHz

 (-3dB) can be examined.

Sinewave signals of 

50MHz

 are displayed with a height of

approx. 50% (-6dB). However when examining square or pulse
type waveforms, attention must be paid to the 

harmonic

content

 of such signals. The repetition frequency (fundamen-

tal frequency) of the signal must therefore be significantly
smaller than the upper limit frequency of the vertical amplifier.
Displaying composite signals can be difficult, especially if they
contain no repetitive higher amplitude content which can be
used for triggering. This is the case with bursts, for instance. To
obtain a well-triggered display in this case, the assistance of the

variable holdoff

 and/or 

variable time control

 may be required.

Television 

video signals

 are relatively easy to trigger using the

built-in 

TV-Sync-Separator (TV). 

For optional operation as a

DC or AC voltage amplifier, the vertical amplifier input is
provided with a 

DC/AC

 switch. The 

DC

 position should only be

used with a series-connected attenuator probe or at very low
frequencies or if the measurement of the DC voltage content
of the signal is absolutely necessary.

When displaying very low frequency pulses, the flat tops may be
sloping with 

AC

 coupling of the vertical amplifier (

AC

 limit

frequency approx. 1.6 Hz for 3dB). In this case, 

DC

 operation is

preferred, provided the signal voltage is not superimposed on a
too high DC level. Otherwise a capacitor of adequate capacitance
must be connected to the input of the vertical amplifier with DC
coupling. This capacitor must have a sufficiently high breakdown
voltage rating. 

DC

 coupling is also recommended for the display

of logic and pulse signals, especially if the pulse duty factor
changes constantly. Otherwise the display will move upwards or
downwards at each change. Pure direct voltages can only be
measured with 

DC

-coupling.

Amplitude Measurements

In general electrical engineering, alternating voltage data
normally refers to effective values (rms = root-mean-square
value). However, for signal magnitudes and voltage designations
in oscilloscope measurements, the peak-to-peak voltage (Vpp)
value is applied. The latter corresponds to the real potential
difference between the most positive and most negative
points of a signal waveform.

If a sinusoidal waveform, displayed on the oscilloscope screen,
is to be converted into an effective (rms) value, the resulting
peak-to-peak value must be divided by 2x

2 = 2.83. Conversely,

it should be observed that sinusoidal voltages indicated in Vrms
(Veff) have 2.83 times the potential difference in V

pp

. The

relationship between the different voltage magnitudes can be
seen from the following figure.

Voltage values of a sine curve

Vrms = effective value; Vp= simple peak or crest value;
Vpp = peak-to-peak value; Vmom = momentary value.

The minimum signal voltage which must be applied to the Y
input for a trace of 1div. height is 

1mVpp

 when the 

Y-MAG. x5

pushbutton is depressed, the 

VOLTS/DIV

. switch is set to

5mV/div., and the vernier is set to 

CAL

 by turning the 

fine

adjustment knob

 of the

 VOLTS/DIV. 

switch fully clockwise.

However, smaller signals than this may also be displayed. The

deflection coefficients

 on the input attenuators are indicated

in 

mV/div.

 or 

V/div.

 (peak-to-peak value).

The magnitude of the applied voltage is ascertained by
multiplying the selected deflection coefficient by the vertical
display height in div.

If an attenuator probe x10 is used, a further multiplication by a
factor of 10 is required to ascertain the correct voltage value.

For exact amplitude measurements, the variable control on the
attenuator switch must be set to its calibrated detent

 CAL.

When turning the variable control ccw, the sensitivity will be
reduced by a factor of  2.5.

Therefore every intermediate value is possible within the 1-2-
5 sequence.

With direct connection to the vertical input, signals 

up to

400Vpp

 may be displayed (attenuator set to 

20V/div.

, variable

control to left stop).

With the designations

H

display height in div.

,

U

signal  voltage in Vpp

 at the vertical input,

D

deflection coefficient in V/div.

 at attenuator switch,

the required value can be calculated from the two given
quantities:

However, these three values are not freely selectable. They
have to be within the following limits (trigger threshold, accuracy
of reading):

H

between 0.5 and 8div., if possible 3.2 to 8div.,

U

between 0.5mVpp and 160Vpp,

D

between 1mV/div. and 20V/div. in 1-2-5 sequence.

Examples:
Set deflection coefficient D = 50mV/div.   0.05V/div.,
observed display height H = 4.6div.,
required voltage U = 0.05·4.6 = 0.23Vpp.

Input voltage U = 5Vpp,
set deflection coefficient D = 1V/div.,
required display height H = 5:1 = 5div.

Signal voltage U = 230Vrms·2

2 = 651Vpp

(voltage > 160Vpp, with probe 10:1: U = 65.1Vpp),
desired display height H = min. 3.2div., max. 8div.,
max. deflection coefficient D = 65.1:3.2 = 20.3V/div.,
min. deflection coefficient D = 65.1:8 = 8.1V/div.,
adjusted deflection coefficient D = 10V/div.

The input voltage must not exceed 400V, independent
from the polarity.

 If an AC voltage which is superimposed on

a DC voltage is applied, the maximum peak value of both
voltages must not or –400V. So for AC voltages with
a mean value of zero volt the maximum peak to peak value
is 800Vpp. If attenuator probes with higher limits are used,
the probes limits are valid only if the oscilloscope is set to DC
input coupling.

Type of signal voltage

Содержание HM303-6

Страница 1: ...O s c i l l o s c o p e H M 3 0 3 6 Manual English...

Страница 2: ...lifiers 12 X Y Operation 13 Phase comparison with Lissajous figures 13 Phase difference measurement 13 in DUAL mode 13 Phase difference measurement in DUAL mode 14 Measurement of an amplitude modulati...

Страница 3: ...effected by such fields Although the interior of the oscilloscope is screened by the cabinet direct radiation can occur via the CRT gap As the bandwidth of each amplifier stage is higher than the tot...

Страница 4: ...oise Amplifiers Time Base 0 2 s 100 ns cm with X Magnification to 10 ns cm Triggering from 0 to 50 MHz from 5 mm signal level 100 MHz 8 mm Analog mode provides unexcelled signal presentation at high r...

Страница 5: ...1 s div 1 2 5 Sequence Accuracy 3 Variabel uncalibrated 2 5 1 to 0 5 s div X Magnification x 10 up to 10 ns div Accuracy 5 Hold Off Time variable to approx 10 1 XY Bandwidth X Amplifier 0 2 5 MHz 3 d...

Страница 6: ...Without pulling the locking knobs they will latch in into the next locking position Handle mounting dismounting The handle can be removed by pulling it out further depending on the instrument model in...

Страница 7: ...e atmosphere or if there is danger of explosion The operating position may be any however sufficient ventilation mustbeensured convectioncooling Prolongedoperationrequires the horizontal or inclined p...

Страница 8: ...83 times the potential difference in Vpp The relationship between the different voltage magnitudes can be seen from the following figure Voltage values of a sine curve Vrms effective value Vp simple p...

Страница 9: ...frequency Depending on the time base setting of the TIME DIV switch one or several signal periods or only a part of a period can be displayed The time coefficients are stated in s div ms div and s div...

Страница 10: ...d The attenuator switch must then be turned back to the left until the vertical signal height is only 3 8div With a signal amplitude greater than 160Vpp an attenuator probe must be inserted before the...

Страница 11: ...be visible after a short warm up period of approx 10 seconds Adjust Y POS I and X POS controls to center the baseline Adjust INTENS intensity and FOCUS controls for medium brightness and optimum sharp...

Страница 12: ...is slightly more difficult but causes a better result The rising edge should be as steep as possible with a pulse top remaining as straight and horizontal as possible After completion of the HF adjus...

Страница 13: ...uite simple with the following formula and a pocket calculator with trigonometric functions Apart from the reading accuracy the signal height has no influence on the result The following must be noted...

Страница 14: ...tion F 1 MHz f 1 kHz m 50 UT 28 3 mVrms If the two values a and b are read from the screen the modulation factor is calculated from where a UT 1 m and b UT 1 m The variable controls for amplitude and...

Страница 15: ...a positive potential That has nothing to do with zero or ground potential and absolute voltage values The positive slope may also lie in a negative part of a signal A falling edge sign triggers when...

Страница 16: ...ossible in this trigger mode External triggering The internal triggering is disconnected by depressing the TRIG EXT button The time base can be triggered externally via the TRIG EXT socket using a 0 3...

Страница 17: ...are required The test leads are connected to the insulated socket and the adjacent ground socket beneath the screen The component can be connected to the test leads either way round After use to retu...

Страница 18: ...can be identified by comparison with a known diode Testing Transistors Three different tests can be made to transistors base emitter base collector and emitter collector The resulting test patterns ar...

Страница 19: ...soldered This terminal should then be connected to the insulated COMP TESTER socket avoiding hum distortion of the test pattern Another way is a test pattern comparison to an identical circuit which i...

Страница 20: ...II display SLOPE Selects the slope of the trigger signal Element Function SLOPE rising edge pushbutton switch falling edge TR LED lights if sweep is triggered LED LEVEL Adjustment of trigger level kno...

Страница 21: ...ast 2 5 fold For time measurements turn to right hand stop Element Function XY Selects X Y operation pushbutton switch stops sweep X signal via CH I Attention Phosphor burn in without signal TRIG EXT...

Страница 22: ...ggering TRIG EXT pushbutton depressed sync signal 0 3Vpp to 3Vpp to TRIG EXT socket Line triggering normal triggering AT NM and ALT pushbuttons depressed Select trigger coupling with TRIG MODE selecto...

Страница 23: ...M A N U A L H A N D B U C H M A N U E L...

Страница 24: ...stra e 6 A Rohde Schwarz Company D 63533 Mainhausen registrierte Marke Tel 49 0 61 82 800 0 DQS Zerti kation DIN EN ISO 9001 2000 Fax 49 0 61 82 800 100 Reg Nr 071040 QM sales hameg de Oscilloscopes S...

Отзывы: