ORCA-Flash4.0 LT3 Digital CMOS camera C11440-42U40 Instruction manual_Ver.1.1
41
(3) Operating environment
Ambient operating temperature
0 ºC to + 40 ºC (Fan Speed: Fast)
0 ºC to + 35 ºC (Fan Speed: Middle, Slow)
Ambient storage temperature
-10 ºC to + 50 ºC
Ambient operating humidity
30 % to 80 %, no condensation
Ambient storage humidity
Less than 90 %, no condensation
Place of operating
Indoor, altitude up to 2000 m
(4) Dimensional outline and weight
Dimensional outline 85 mm
(W)
× 85.5 mm
(H)
× 117.6 mm
(D)
Weight
Camera
Approx. 1.1 kg
AC a power supply cord
Approx. 1.0 kg
Note
•
Please see 14. [DIMENSIONAL OUTLINES] for detail of dimensions.
(5) Applicable standards
EMC
EN61326-1: 2013 Class A
Emission limits: CISPR 11 Group1 Class A
Immunity requirements: Table2
Function performance and operation
mode in immunity tests
Intensity fluctuation of image is within ±5 %.
Restoration procedure in immunity
tests
Exit the image acquisition software, turn off the
power of camera and focus adjustment unit.
And then turn on the power of each device.
Open the image acquisition software.
FCC
47 CFR FCC Part15 Subpart B Class A
Содержание ORCA-Flash4.0 LT3
Страница 2: ...Blank page...