12 

 

5. 

EQ-400 Operation  

5.1.  Startup 

Once the lamp is set up properly, verify that all personnel that will be in contact with the lamp system are 
aware of the potential hazards involved.  It is the responsibility of the user to verify that the lamp is being used 
safely and to consult their laser safety officer for safety measures specific to their use of the source. All safety 
measures appropriate for a Class 4 laser product should be taken, including protective eyewear as described 
in Section 1. 

1.  Verify that all interlocks are in place and none of the red fault LEDs on the  Controller front panel 

are lit.  The green POWER OK LED should be lit. 

2.  Place the keyswitch in the ENABLE position. 

3.  Turn on the OPERATE Switch (place switch in UP position). 

Note:

 If the lamp has been on within the last 5 minutes, the POWER OK LED will be flashing and 

re-start will not be possible.  This is to allow the Lamp to cool down before the next ignition cycle 

4.  Within several seconds the LASER ON LED will light.  Laser light is now present in the Lamp.   

5.  Within several seconds the igniter will be turned on automatically and the plasma will ignite.  The 

LAMP ON LED will light. To avoid thermal shock to the Lamp components, the EQ-400 starts with 
the laser at half power.  After 20 seconds the laser will switch to full power.  

5.2.  Stopping 

The following steps explain the proper procedures stopping  
 
To turn the lamp off, simply turn the OPERATE Switch to the OFF position.  It is recommended to turn the 
keyswitch to the DISABLE position. The key can be removed if needed for safety or personnel access 
reasons.  If the lamp will not be used for some time, the AC supply can be turned off. 

To minimize wear on the ignition components, it is best to avoid frequently starting and stopping the lamp.  It 
is recommended to run the lamp continuously if long off-periods are not required. 

5.3.  Manual Reset 

If any fault occurs during operation of the lamp (including actuation of the remote interlock), the laser is 
immediately shut off, and light output ceases. When the fault condition is cleared, the lamp will not 
automatically restart.  The system must be manually reset by turning the OPERATE switch to the OFF 
position.  Then, assuming that the fault or interlock condition no longer exists, the lamp can be restarted by 
turning the switch to the OPERATE position.  

If a fault occurs while the lamp is not operating (switch in the OFF position), the switch must be turned to 
OPERATE and then back to OFF to clear the fault.  The lamp can then be started normally by turning the 
switch to OPERATE. 

 

6. 

Service Requirements 

There are no user-serviceable parts inside the EQ-400.  For any problems encountered during operation, 
please contact Energetiq Technology for assistance.   

Содержание ENERGETIQ LDLS EQ-400

Страница 1: ...Model EQ 400 Laser Driven Light Source LDLS Operation Manual Revision 9 May 2022 Part Number DOC 7388...

Страница 2: ...ical information including drawings schematics and specifications contained in this document are the property of Energetiq and shall not be reproduced in whole or in part without the written consent o...

Страница 3: ......

Страница 4: ...Remote Interface 10 3 6 External interlock connection 10 4 EQ 400 Setup 10 4 1 Connections 10 4 2 Chiller Information 11 5 EQ 400 Operation 12 5 1 Startup 12 5 2 Stopping 12 5 3 Manual Reset 12 6 Serv...

Страница 5: ...Section 4 for connection details For additional information contact Energetiq Technology Technical Support Services at 1 781 939 0763 X111 or e mail EUVproductsupport Energetiq com 2 Safety instructi...

Страница 6: ...not open the enclosure of either the Lamp or the Controller Dangerous invisible infrared laser beams and hazardous voltages exist inside the Lamp Opening the chassis both voids the warranty and expose...

Страница 7: ...on 400W 1070nm pump laser under a failure mode condition where aperture is itself accessible 4 meters Required Optical Density OD for safety eyewear 6 or higher at 1070 nm These values are based on a...

Страница 8: ...and Safety Notification The following safety labels appear on the product Table 1 shows the location of each label on the EQ 400 system and Table 1 describes their meanings Figure 1 EQ 400 Safety lab...

Страница 9: ...e outlet aperture for laser radiation 4 UV Hazard warning label indicates hazardous levels of UV light are present 5 Protective housing label warns of an enclosed laser radiation hazard 6 Explanatory...

Страница 10: ...00 The shipping box should contain EQ 400 Power Supply Controller unit EQ 400 Lamp unit Accessory Box Interconnecting cable from Lamp to Power Supply Controller 9 pin D sub AC Power cable AC Power plu...

Страница 11: ...dicators are described below in Table LED Label Meaning when lit POWER OK AC power is connected to the EQ 400 PCU LAMP ON UV Light is on LASER ON Laser power is ON and laser light is being produced wi...

Страница 12: ...unctions This connector also functions as the remote interlock connector 4 RS 485 Connector 9 pin D sub Connector for optional RS 485 interface See Sections 3 and 4 for electrical details and commands...

Страница 13: ...mp 3 Control Connector 9 pin D sub Provides various power and control signals to from the Power Supply Controller No other connector or cable may be used with the EQ 400 other than the one supplied 4...

Страница 14: ...he Controller weighs 18 8 kg 41 5 lb A two person lift is recommended Use care when moving the Controller not to bend the fiber optic cable sharply Minimum bend radius is 50mm 2 Connect the Lamp unit...

Страница 15: ...rlock switch to the interlock pins of the remote I O connector functions as remote interlock connector If using remote control mode connect user s control system to the remote I O connector using a su...

Страница 16: ...conds the laser will switch to full power 5 2 Stopping The following steps explain the proper procedures stopping To turn the lamp off simply turn the OPERATE Switch to the OFF position It is recommen...

Страница 17: ...rgetiq if condition persists LAMPHOUSE OVERTEMP Lamphouse internal temperature too high 1 Check for sufficient lamphouse cooling water flow 2 Ensure that cooling water temperature is between 18 and 24...

Страница 18: ...resentative SMC PN HRS018 A 10 T PN Description EQ 400 CHILLER KIT EQ 400 Chiller with Tubing Kit Note Cooling fluid is not included and must be locally sourced 9 Warranty Energetiq Technology Inc s w...

Страница 19: ...15 EQ 400 Controller top in mm 10 1 2 Lamp Dimensional Drawings EQ 400 Side view in mm...

Страница 20: ...y non condensing 80 max for temperatures up to 31 C decreasing linearly to 50 max at 35 C Pollution Degree 2 normally only non conductive pollution occasional temporary condensation possible Installat...

Страница 21: ...n when ON LAMP MODULE FAULT 3 Pulled to digital common when OK float on FAULT CONTROLLER FAULT 4 Pulled to digital common when OK float on FAULT ISOLATED 5V SUPPLY 5 200mA maximum referenced to digita...

Страница 22: ...face Commands The RS 485 serial interface is provided for user adjustment of laser operating power to increase or decrease light output The EQ 400 is shipped with the laser power set to a factory defa...

Страница 23: ...actory default value Power XXX X B Saves present settings to flash memory This command should be issued after the settings are at their desired values If not changes will be lost if power to the EQ 40...

Страница 24: ...1 out of range 0 OK 5 Laser current source 1 fault 0 OK 6 Lamphead temperature 1 overtemperature 0 OK 7 PCU temperature 1 overtemperature 0 OK 8 Lamphead interlock 1 open 0 OK 9 External interlock 1...

Страница 25: ...21 Energetiq Technology Inc 205 Lowell St Wilmington MA 01887 Phone 1 781 939 0763 Fax 1 781 939 0769 Email info energetiq com www energetiq com DOC 7388 Rev 9 5 22...

Отзывы: