background image

 

 

 

 

KXX-A12040 A

 

Never  connect  the  product  to  the  PC  before  installing  the  evaluation  software  into  the  PC.  This  might 
prevent correctly installing the evaluation software. 

 

The  supported  OS  is  Microsoft

®

  Windows

®

.  Refer  to  the  datasheet  or  operation  manual  for  details  on 

versions and platforms. 

 

Depending on the PC model, the power-saving mode or screensaver might interrupt power supplied from 
the USB port. Interruption of power supplied from the USB port will also stop the product operation and 
might cause problems when power is later resupplied from the PC after recovering from power-saving or 
screensaver  mode.  If  the  PC  causes  this  problem,  then  please  disable  the  power-saving  mode  and 
screensaver. (See the PC operation manual to find PC functions and settings.) 

 
(1) TG-CCD/TM-CCD series 

An external power supply is required to operate the internal image sensor. Use the AC adapter that 
comes supplied with the product. 

 
(2) TG-cooled series 

 

An external power supply is required to operate the internal thermoelectric cooler and cooling fan. Please 
prepare a low noise power supply and connect it to the product using the connector that came supplied with 
it. Use caution when turning on the power supply because an inrush current flows at that time. Table 1 
shows the maximum current in steady state and typical inrush current. 

 
[Table 1] Maximum current in steady state and typical inrush current (TG-cooled series) 

Power supply 

Mini-spectrometer type 

Supply 

voltage 

(V) 

Maximum current in 

steady state 

(A) 

Typical inrush 

current 

(A) 

For thermoelectric 

cooler 

C9913GC 

1.8 

C9914GB 

C11118GA 

2.8 

For cooling fan 

All models 

12 

0.2 

0.4 

[Power supply examples] 
Multi-output power supply: PW18-1.3ATS (Texio Technology Corporation) 
Unit type power supply (+5 V): NNS15-5 (TDK-Lambda Corporation) 

(+12 V): PBA10F12-N (Cosel Co., Ltd.) 

 

 

When the cooling fan filter becomes dirty, replace it by following the procedure described in the product 
operation manual. 

 

Do not block the air intake and exhaust ports for the cooling fan during use. If the temperature inside the 
product rises too high, the internal safety circuit acts to stop product operation. However, high temperatures 
may cause a malfunction. If product operation stopped due to the safety circuit, then immediately turn off 
the  external  power  supply  and  disconnect  this  product  from  the  USB  port  of  the  PC.  Then,  check  and 
eliminate the cause of the problem. Resume operation after making sure that the product temperature has 
dropped to room temperature. 

Содержание C12666MA

Страница 1: ...when no light is input the product generates an output consisting of an offset component and dark signal The offset component is not dependent on integration time but the dark signal is dependent on...

Страница 2: ...e supplied with it Use caution when turning on the power supply because an inrush current flows at that time Table 1 shows the maximum current in steady state and typical inrush current Table 1 Maximu...

Страница 3: ...h a wrist strap having protective resistance of 750 k to 35 M directly to the skin and ground the strap If the wrist strap does not include protective resistance there is a risk of electric shock haza...

Страница 4: ...ductive case Avoid using plastic or styrofoam as they may generate static electricity by vibration during shipping etc causing breakdown or deterioration of the heed type Use a conductive carrying cas...

Страница 5: ...fluctuations so separate it as far as possible from the video signal line and voltage supply line Apply the input clock pulses at the specific timing and voltage to the product connection terminals on...

Страница 6: ...the optical fiber SMA connector and then use the slotted holes to secure the mini spectrometer at a position where the optical fiber is free from stress Figure 3 Precautions when connecting input opt...

Страница 7: ...cure it with resin or a holder etc 3 Precautions for designing the circuit board The packaging of the micro spectrometer is electrically conductive so be careful when designing the circuit to avoid sh...

Страница 8: ...d of time seal it in a moisture proof package containing silica gel or keep it in a low humidity desiccator 4 Input optical fibers option When connecting the input optical fiber clean the ferrule on t...

Отзывы: