External trigger mode
The exposure time can be set independently for the top and bottom areas with Edge trigger
mode and Start trigger mode in W-VIEW Mode. The exposure time for both halves is the same
with Level trigger mode and Synchronous readout trigger mode.
Figure 9-21
(Edge trigger mode)