35
9-3 W-VIEW MODE
9-3-1
Readout method (scan mode)
The readout direction can be set for the top area and the bottom area independently in W-VIEW
Mode. (Figure 9-16, 17, 18, 19)
Figure 9-16
Top: top to center
Figure 9-17
Top: center to top
/ Bottom: center to bottom
/ Bottom: bottom to center
Figure 9-18
Top: top to center
Figure 9-19
Top: center to top
/ Bottom: bottom to center
/ Bottom: center to bottom
W-VIEW Mode supports Binning readout, Sub-array readout and Rapid rolling mode.
In sub-array mode, the vertical position of sub-array readout can be set independently in the top
and bottom areas.
The vertical height of sub-array readout is same in the both areas.
Note
•
Please refer to 9-3-3 [Frame rate calculation] about the frame rate of each readout mode.
Rapid Rolling mode is different from other modes.