  TRIGGER OUTPUT 

9-8

The camera provides a range of trigger output signals to synchronize with an external instrument 

and the camera becomes the master and the external instrument becomes the slave. There are 

three different trigger output functions as follows. 

 

-  Global exposure timing output 

-  Programmable timing output 

-  Trigger ready output   

 

Also, it can output continuous High output (High output fixed) or continuous Low output (Low 

output fixed). 

These trigger output functions can be selected by software command, and they are output from 

Timing out connector. And the polarity can be changed by software command. 

 

9-8-1  GLOBAL EXPOSURE TIMING OUTPUT 

Global exposure timing output shows the period of all pixels exposing.   

 

 

• 

For more detail, please refer to the timing charts of Figure 9-1 to 9-8.   

 

9-8-2  PROGRAMMABLE TIMING OUTPUT 

By using the programmable timing output, synchronizing external devices is simple. A system 

that needs simple timing signal does not require a delay unit nor pulse generator. It is possible to 

program and output a pulse that has an optional pulse width and an optional delay time to the 

exposure start timing of the camera by command. 

The setting range for delay time is 0 µs to 10 s, the setting range for pulse width is 1 µs to 10 s, and 

the minimum setting step is respectably 1 µs.   

 

 

Figure 9-9  Programmable timing output (Edge trigger mode, Rising edge) 

 

*1  Delay + Jitter = 

3H (≈ 37.3 µs) + up to 1H (≈ 12.4 µs)

 

Internal exposure time 

setting

Exposure start signal

0H

1H

2H

(LAST-1)H

(LAST)H

Sensor readout

Camera data output

Programmable timing 

output

External trigger

Exposure

Delay

Width

Exposure start timing

*1

Содержание C11440-36U

Страница 1: ...and damage to property when using this camera Be sure to read this Instruction manual beforehand in order to use the digital camera correctly The manual describes the correct method of handing the ca...

Страница 2: ...Blank page...

Страница 3: ...eath Improper handling of the camera without observing these cautions could lead to personal injury to the user or damage to property Note This symbol indicates a note to help you get the best perform...

Страница 4: ...electric shock Do not attempt to dismantle or modify the camera Doing so can also lead to damage and even injury as some internal components become very hot Only touch parts as indicated in this manua...

Страница 5: ...C mount of the camera Screwing it in excessively can damage the camera Shipping precautions When transporting the camera by truck ship airplane etc wrap it securely in packaging material or something...

Страница 6: ...pply cord for AC adapter 1 Lens mount cap attached to the camera 1 C11440 36U Before Use Booklet 1 C11440 36U Instruction manual CD ROM 1 Option SMA BNC cable A12106 05 SMA SMA cable A12107 05 USB 3 0...

Страница 7: ...dew condensation might generate Close to a strong source of magnetism or radio waves Where there is vibration Where it might come into contact with corrosive gases such as chlorine or fluorine Where...

Страница 8: ...e 16 9 3 2 Electrical shutter mode 16 EXTERNAL TRIGGER MODE 17 9 4 9 4 1 Edge trigger mode 17 9 4 2 Level trigger mode 18 9 4 3 Synchronous readout trigger mode 18 9 4 4 Start trigger mode 20 9 4 5 Tr...

Страница 9: ...C11440 36U Instruction Manual Ver 1 1 7 CAMERA SPECIFICATIONS 27 13 1 14 DIMENSIONAL OUTLINES 30 15 WARRANTY 31 16 CONTACT INFORMATION 32...

Страница 10: ...pixel amplifiers is greatly minimized In addition there is the on chip CDS correlated double sampling circuit which plays an important role in achieving low noise Moreover the data of one horizontal l...

Страница 11: ...is performed in real time without scarifying the readout speed at all 7 USB 3 0 bus Interface Image signals undergo A D conversion inside the camera and are transmitted externally via the USB 3 0 bus...

Страница 12: ...rew the lens more than 5 mm onto the C mount of the camera Screwing it in excessively can damage the camera STATUS lamp STATUS The LED indicates status of camera Lighting color Status of power distrib...

Страница 13: ...ing operated using external synchronization Input is 3 3 V LVCMOS level and input impedance is 10 k When External trigger is input the trigger is activated at the falling or rising edge of the signal...

Страница 14: ...r of the computer corresponding to USB 3 0 When the USB 3 0 cable is connected the application software must be closed and the camera must be turned off Hamamatsu recommends A12046 03 optional USB 3 0...

Страница 15: ...art operating the camera 1 Connect the cable as shown in Figure 7 1 before you start operation 2 Turn on the power of computer END OF IMAGING 8 2 Follow the procedure as below when imaging is finished...

Страница 16: ...tant role in achieving low noise In addition the data of one horizontal line is read by the on chip column amplifier and A D in parallel and simultaneously Therefore the CMOS image sensor realizes bot...

Страница 17: ...s possible to increase the frame rate by reducing the number of lines scanned The frame rate is determined only by vertical width the horizontal width does not affect to the frame rate The Sub array s...

Страница 18: ...dout mode is used so that the exposure time can be extended to collect more signals and increase the signal to noise ratio if the object is dark Figure 9 2 Normal readout mode 9 3 2 ELECTRICAL SHUTTER...

Страница 19: ...used so that the exposure starts according to an external signal Exposure time is set by software command In this mode the exposure begins on the edge rising falling edge timing of the input trigger s...

Страница 20: ...me 9 4 3 SYNCHRONOUS READOUT TRIGGER MODE Synchronous readout trigger mode is used for continuous imaging when it is necessary to control the exposure start timing of each frame from an external sourc...

Страница 21: ...trigger mode Rising edge 2 Pulse count Also in Synchronous readout trigger mode synchronous readout can be controlled by specifying set by command the number of timing pulses to determine the exposur...

Страница 22: ...for delay time is 0 s to 10 s and the minimum setting steps is 1 s SOFTWARE TRIGGER MODE 9 5 The camera has two software trigger functions Edge trigger mode and Start trigger mode Each software trigg...

Страница 23: ...ber 1 to 803 572 1H 1 37 125 MHz 462 clk 12 4 s In the case of Level trigger and Synchronous readout trigger mode jitter which is up to 1H 12 4 s might be included in exposure time FRAME RATE CALCULAT...

Страница 24: ...period of all pixels exposing For more detail please refer to the timing charts of Figure 9 1 to 9 8 9 8 2 PROGRAMMABLE TIMING OUTPUT By using the programmable timing output synchronizing external dev...

Страница 25: ...hows the trigger ready period when the camera can accept an external trigger in External trigger mode For more detail please refer to the timing charts of Figure 9 1 to 9 4 ANALOG GAIN FUNCTION 9 9 Th...

Страница 26: ...erate white spot After an image made using a certain exposure time is loaded the CMOS image sensor is exposed to darkness for the same amount of time and another image is obtained After this the diffe...

Страница 27: ...the window gently Use Lens Cleaning Paper for cleaning of glass window in front of the image sensor Please use a plastic tweezers and take extra care not to scratch the glass window with the tweezers...

Страница 28: ...n Cause Measures Chapter Lens is dirty Wipe the lens 2 Image is blurred Cause Measures Chapter Lens is not focused Contact Hamamatsu subsidiary or local distributor 16 CMOS image sensor is dirty 3 Onl...

Страница 29: ...capacity and the readout noise 2 Camera functions specifications a Image readout Readout mode Full resolution Digital binning 2 Sub array Full resolution Number of pixel 1920 H 1200 V Maximum frame ra...

Страница 30: ...nous readout 0 to 4095 times Trigger delay function 0 s to 10 s 1 s step Software trigger Trigger input mode Edge Start Exposure time 26 17 s to 10 s Trigger delay function 0 s to 10 s 1 s step c Trig...

Страница 31: ...0 C to 50 C Ambient operating humidity 30 to 80 no condensation Ambient storage humidity 90 no condensation 5 Dimensional outline and weight Dimensional outline 80 mm W 80 mm H 51 5 mm D Weight of the...

Страница 32: ...C11440 36U Instruction Manual Ver 1 1 30 14 DIMENSIONAL OUTLINES Figure 14 1 Dimensional outlines Front Side Rear Under Unit mm...

Страница 33: ...period in the event of a natural disaster or if you handle the system contrary to the instructions in this manual use it without due caution or try to modify it 3 We will repair the system or replace...

Страница 34: ...be found under www hamamatsu com The contents of this manual are subject to change without notice The unauthorized reproduction or distribution of parts or all of this manual is prohibited If one of...

Отзывы: