7
Figure 5
Figure 6
Figure 7
4. When all air has been removed from the entire tubing and
fluid is observed at end of luer lock, the administration set
is primed.
5. Turn the dial back to
ø
, off position and close the clamp
(Figure 6).
6. Replace tubing cap until ready for use.
Starting Infusion
Use Aseptic Technique
1. Connect tubing to patient’s catheter. Make sure
connection is secure.
2. Select the appropriate flow rate by turning the dial on
the SELECT-A-FLOW* device until the flow rate setting is
within the window and aligned with the ml/hr
x
mark on
the face of the SELECT-A-FLOW* device (Figure 7, 8).
• The tactile feel will allow the user to ensure selected
flow rate is set.
3. Open clamp.
Figure 8
Examples of flow rate settings
3 ml/hr
6 ml/hr
Caution:
Flow rate setting must be within window and
aligned with ml/hr
▼
mark to ensure accurate flow rate. Do not
dial between numbers. Flow rate is unpredictable if dialed between
numbers (Figure 9).
Figure 9
Correct alignment
Incorrect alignment
Содержание ON-Q Select-A-Flow
Страница 1: ...PUMP WITH SELECT A FLOW VARIABLE RATE CONTROLLER Instructions for Use...
Страница 2: ......
Страница 4: ...2...
Страница 46: ...44 PVC 2 DEHP DEHP DEHP DEHP PVC 1 Halyard Health...
Страница 94: ...92 2 1 Halyard Health...
Страница 98: ...96 3 5 6 7 4 5 6 6 1 2 SELECT A FLOW SELECT A FLOW 7 8 3 8 3 6 9 9...
Страница 123: ...121 1 DEHP PVC DEHP DEHP DEHP PVC 1 Halyard Health SELECT A FLOW...
Страница 130: ...128 DEHP 2 PVC DEHP DEHP DEHP PVC 1 Halyard Health...
Страница 140: ...138 5 6 7 4 5 6 6 1 2 SELECT A FLOW SELECT A FLOW ml 7 8 3 8 3 ml 6 ml ml 9 9 10 10...
Страница 142: ...140...
Страница 143: ...141...