halstrup-walcher PS27 Скачать руководство пользователя страница 8

 

Operating Instructions PS27 

 

3.3 Connections for Two-Wire System (Option): 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
One measurement range   

 

 

Four measurement ranges 

 

ST4 Pin 

Connection name 

Meaning 

24V 

+terminal 15-32VDC 

GND 

Not used 

OUT 

-terminal 

GND 

Not used 

SET 0 

Zero-point input 

 

3.4 Connections M12-plug 

 

 

Pin 

Connection name 

24 VDC 

Output signal 

GND 

GND output signal 

 

4. 

Resetting Offset 

After starting the pressure transmitter, give it some 15 minutes to warm up. 
During this time the output signal may be unstable. 
 

Zero-point calibration of the appliance is only effected via the input assigned for this 
purpose. 

The user must ensure that the appliance is not subjected to any 

pressure during the calibration process as this might lead to a wrong offset 
value which, in turn, would result in a wrong output signal. 

The 24V input makes 

it possible to activate a 3/2 directional control valve with the same signal, which 
switches the PS27's pressure inputs at zero pressure during the zero calibration 
process. In this case the zero point calibration signal must be applied until there is no 
doubt that the pressure has fallen to zero.  

 

While Resetting Offset is in progress, the power supply must not be 

switched off. This might result in a loss of the device setting and may 

leave the PS27 unusable. 

5  4  3  2  1  

5  4  3  2  1  

Содержание PS27

Страница 1: ...ential Pressure Transmitter PS27 halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E mail info halstrup walcher com Internet www hal...

Страница 2: ...Code 2022 The manufacturer owns the copyright to this instruction manual It contains technical data instructions and drawings detailing the device s features and how to use them It must not be copied...

Страница 3: ...must be given to these operating instructions and in particular to the associated safety instructions Do contact the manufacturer should you not understand all the sections Be heedful of the operatin...

Страница 4: ...er 6 3 Startup 6 3 1 Functional Description 6 3 2 Connections for Four Wire version standard in the device 7 3 3 Connections for Two Wire System Option 8 3 4 Connections M12 plug 8 4 Resetting Offset...

Страница 5: ...ator Only instructed persons as assigned by the plant operator are to operate the transmitter Do not perform any functional test involving compressed or breathable air This would otherwise damage thos...

Страница 6: ...ional Description PS27 pressure transmitters are of the precision type and despite their sturdiness they should still be treated with care Installation right next to heat and radiation sources e g rad...

Страница 7: ...gram on the appliance cover circuit board mount ST4 Pin Connection name Meaning 1 24V Power supply input 24VAC DC 2 GND Ground for power supply 3 OUT Voltage or current output 4 GND Ground for output...

Страница 8: ...ed via the input assigned for this purpose The user must ensure that the appliance is not subjected to any pressure during the calibration process as this might lead to a wrong offset value which in t...

Страница 9: ...ching of 4 measuring ranges into the appliance has been undertaken by the manufacturer The switches 1 and 2 of SW can be used to select the measuring range in question before the transmitter is switch...

Страница 10: ...sor s output so that any adjusted time delay will also cause a delay in relay triggering 9 In case of Malfunctions Fault description Possible cause Corrective action No output signal Supply voltage is...

Страница 11: ...o Fitting 4 or 6mm Output signals 0 2 10 V RL 50 kOhm 0 4 20 mA RL 500Ohm 4 20 mA two wire 17 32 V RL 500Ohm Switching contact optional max 230 VAC 6 A min 500 mW Display optional 3 digit LCD Overpres...