Specifications 

 

Measurement accuracy: 

± 2 % of the set value for 

≥ 100 Pa 

± 3 % of the set value for < 100 Pa 

 
Temp. coefficient zero point: 

± 0.1 % FS / K 

Temp. coefficient span: 

± 0.1 % FS / K 

 
Measurement ranges: 

50 / 100 / 200 / 500 Pa 
1 / 2 / 5 / 10 kPa 
  

 

As an option, the customer can use a DIP-switch to 
select 4 pre-set measuring ranges 

 

Step response time (T63): 

20 ms

…4s (adjustable factory provided) 

(Time constant) 
 
Pressure ports: 

Nominal 4 or 6 mm width for hose 
Festo-Fitting 4 or 6mm 
 

Output signals: 

0/2…10 V (R

> 50 kOhm) 

 

0/4…20 mA (R

L

 

≤ 500Ohm) 

 

4…20 mA (two-wire 17…32 V, R

L

 

≤ 500Ohm) 

 
Switching contact (optional): 

max. 230 VAC, 6 A; min. 500 mW 

 
Display (optional): 

3½-digit LCD  

 
Overpressure limit: 

Range

 <    2 kPa :    50 kPa 

 

Range > 10 kPa :  200 kPa

 

 
Storage temperature: 

 -20°C...+70°C 

 
Rated temperature range: 

-20°C...+60°C

, with Display 0°C…50°C 

 
Supply voltage: 

24 VDC/VAC ±10 % (no galvanic isolation) 

 

15 … 32 VDC (two-wire) (no relay output) 

 
Power consumption: 

approx. 1VA 

 
Operating position: 

vertical preference 

 
Housing: 

see drawing 

 
Protection class: 

IP 65 

 
Installation: 

The appliance can be installed either on a top-hat rail 
or on a wall with fastening holes 

 
Certificates: 

CE / UKCA 

 

 

Содержание PS27

Страница 1: ...ential Pressure Transmitter PS27 halstrup walcher GmbH Stegener Stra e 10 D 79199 Kirchzarten Germany Phone 49 0 76 61 39 63 0 Fax 49 0 76 61 39 63 99 E mail info halstrup walcher com Internet www hal...

Страница 2: ...Code 2022 The manufacturer owns the copyright to this instruction manual It contains technical data instructions and drawings detailing the device s features and how to use them It must not be copied...

Страница 3: ...must be given to these operating instructions and in particular to the associated safety instructions Do contact the manufacturer should you not understand all the sections Be heedful of the operatin...

Страница 4: ...er 6 3 Startup 6 3 1 Functional Description 6 3 2 Connections for Four Wire version standard in the device 7 3 3 Connections for Two Wire System Option 8 3 4 Connections M12 plug 8 4 Resetting Offset...

Страница 5: ...ator Only instructed persons as assigned by the plant operator are to operate the transmitter Do not perform any functional test involving compressed or breathable air This would otherwise damage thos...

Страница 6: ...ional Description PS27 pressure transmitters are of the precision type and despite their sturdiness they should still be treated with care Installation right next to heat and radiation sources e g rad...

Страница 7: ...gram on the appliance cover circuit board mount ST4 Pin Connection name Meaning 1 24V Power supply input 24VAC DC 2 GND Ground for power supply 3 OUT Voltage or current output 4 GND Ground for output...

Страница 8: ...ed via the input assigned for this purpose The user must ensure that the appliance is not subjected to any pressure during the calibration process as this might lead to a wrong offset value which in t...

Страница 9: ...ching of 4 measuring ranges into the appliance has been undertaken by the manufacturer The switches 1 and 2 of SW can be used to select the measuring range in question before the transmitter is switch...

Страница 10: ...sor s output so that any adjusted time delay will also cause a delay in relay triggering 9 In case of Malfunctions Fault description Possible cause Corrective action No output signal Supply voltage is...

Страница 11: ...o Fitting 4 or 6mm Output signals 0 2 10 V RL 50 kOhm 0 4 20 mA RL 500Ohm 4 20 mA two wire 17 32 V RL 500Ohm Switching contact optional max 230 VAC 6 A min 500 mW Display optional 3 digit LCD Overpres...

Страница 12: ...PS27 12 11 Dimensional Drawing For more information on our measurement technology products please visit our homepage at www halstrup walcher de en products measurement technology 7100 004354L_PS27 doc...

Страница 13: ...Operating Instructions PS27 13 12 Certificates of Conformity...

Страница 14: ...Operating Instructions PS27 14...

Страница 15: ...Operating Instructions PS27 15 Notes...

Отзывы: