A: Calibrating the Wavelength of Jaz
54
013-RD000-000-02-201510
7. Record the Intercept, as well as the First, Second, and Third Coefficients. Additionally, look at
the value for R squared. It should be very close to 1. If not, you have most likely assigned one of
your wavelengths incorrectly.
Keep these values at hand.
Saving the New Calibration Coefficients: USB
Mode
Ocean Optics programs wavelength calibration coefficients unique to each Jaz spectrometer module onto
an EEPROM memory chip in Jaz.
You can overwrite old calibration coefficients on the EEPROM
if you are using Jaz via the USB port.
►
Procedure
To save wavelength calibration coefficients using the USB mode, perform the following steps:
1. Ensure that Jaz is connected to the computer and that you have closed all other applications.
2. Point your browser to
http://www.oceanoptics.com/technical/softwaredownloads.asp
scroll down to
Microcode
. Select
USB EEPROM Programmer
.
3. Save the setup file to your computer.
4. Run the
Setup.exe
file to install the software. The
Welcome
screen appears.
5. Click the
Next
button. The
Destination Location
screen appears.
6. Accept the default installation location, or click the
Browse
button to specify a directory. Then,
click the
Next
button. The
Program Manager Group
screen appears.
7. Click the
Next
button. The
Start Installation
screen appears.
8. Click the
Next
button to begin the installation. Once the installation finishes, the
Installation
Complete
screen appears.
9. Click the
Finish
button and reboot the computer when prompted.
10. Navigate to the
USB EEPROM Programmer
from the Start menu and run the software.
11. Click on the desired Jaz device displayed in the left pane of the
USB Programmer
screen.
12. Double-click on each of the calibration coefficients displayed in the right pane of the USB
Programmer screen and enter the new values acquired in Steps 5 and 6 of the
Wavelength of the Spectrometer
13. Repeat Step 12 for all of the new values.
14. Click on the
Save All Values
button to save the information, and then
Exit
the USB Programmer
software.
The new wavelength calibration coefficients are now loaded onto the EEPROM memory chip on the Jaz
spectrometer module
Содержание Ocean Optics JAZ
Страница 6: ...Table of Contents iv 013 RD000 000 02 201510 ...
Страница 10: ...About This Manual viii 013 RD000 000 02 201510 ...
Страница 18: ...1 Introduction 8 013 RD000 000 02 201510 ...
Страница 26: ...3 Operating Jaz 16 013 RD000 000 02 201510 Jaz Main Menu for DPUs dated January 1 2008 ...
Страница 27: ...4 Operating Jaz 013 RD000 000 02 201510 17 Jaz Main Menu for DPUs dated January 1 2008 Continued ...
Страница 28: ...3 Operating Jaz 18 013 RD000 000 02 201510 Jaz Main Menu for DPUs dated after January 1 2008 ...
Страница 31: ...4 Operating Jaz 013 RD000 000 02 201510 21 for DPUs after January 1 2008 ...
Страница 35: ...4 Operating the JAZ PX Module 013 RD000 000 02 201510 25 JAZ PX Menu JAZ PX Menu Tree ...
Страница 36: ...4 Operating the JAZ PX Module 26 013 RD000 000 02 201510 ...
Страница 44: ...4 Operating the JAZ PX Module 34 013 RD000 000 02 201510 ...
Страница 46: ...5 Operating the JAZ INDY Module 36 013 RD000 000 02 201510 JAZ INDY Menu ...
Страница 47: ...5 Operating the JAZ INDY Module 013 RD000 000 02 201510 37 ...
Страница 48: ...5 Operating the JAZ INDY Module 38 013 RD000 000 02 201510 ...
Страница 49: ...5 Operating the JAZ INDY Module 013 RD000 000 02 201510 39 ...
Страница 54: ...5 Operating the JAZ INDY Module 44 013 RD000 000 02 201510 ...
Страница 60: ...6 Maintenance 50 013 RD000 000 02 201510 ...
Страница 68: ...B Specifications 58 013 RD000 000 02 201510 ...