Get Started:
This box includes:
• Alicat device
• NIST traceable calibration certificate
• Optional accessories
(cables, power adapters, filters)
Power
The power supply
input is on top. Most
devices require
12–24Vdc and 40mA.
See
Power and Signal
Connections
in the
manual for details.
You’re Ready to Connect:
Units
Ensure the proper
units are selected.
Press the button next
to the parameter you
want to change on the
main screen, then
SET
DEVICE ENG UNITS
.
5⁄64
″ 8-32 Nylon-tipped
hex bleed screw
Loosen to bleed
DO NOT REMOVE
Connect the device to your process, and start the
flow of water. To remove air bubbles:
(a) start the flow of liquid,
(b) gently loosen the upstream bleed port screw,
but stop when pressure forces liquid to leak,
(c) tap the flow body.
Do not use bleed ports if the device is inverted.
Ensure the flow path is the same direction as the
flow arrow shown on your unit. For applications
that may continuously introduce air bubbles,
the device may be mounted upside-down.
DOC-MANUAL-L-QUICK-EU
Rev 0 • 2020-07-28
Tare
Before installing
the device into your
process, press the
TARE FLOW
button
on the main screen.
This will zero out
(tare) your unit.
5
Download the full manual:
Alicat.com/manuals
Annual Calibration • Troubleshooting • Questions:
+31 (0) 26 203.1651 • [email protected]