16
Maintenance and Warranty
Maintenance
Cleaning
This device requires minimal maintenance. If necessary, the outside of
the device may be cleaned with a soft dry cloth. Avoid excess moisture
or solvents.
CODA KM-Series mass flow meters used with gas
require no periodic
cleaning, provided they have been flowing clean, dry gas.
CODA KM-Series mass flow meters used with liquids
require some
precautions to avoid contamination and/or corrosion damage. Liquid
should be filtered for particulates or biological materials that may grow
in the device. When removing these units from the line for any extended
period of time, make an effort to remove all of the liquid from the device,
as deposits of calcium or other soluble minerals can affect the accuracy
of the device.
Recalibration
CODA instruments are calibrated to NIST traceable standards at the
time of manufacture. Due to the Coriolis technology, there is not a
factory-recommended periodic recalibration cycle. Recalibration can
be requested at the user’s discretion/requirement by submitting a form
with the device serial number at
.
Replacement Accessories
Accessories are available through support (
), or on our website
at
.
For repair, recalibration, or recycling of this
product contact us (
).
Technical Specifications and Dimensional Drawings
Please visit
to find complete operating specifications
and dimensional drawings.