Preferences:
The preferences page allows users to select the color theme of the HPC301. The
available colors are Gray, Red, Green, and Blue.
The preferences page also allows the user to select the displayed unit of temperature.
The choices are °C (Celsius) and °F (Fahrenheit).
Alarm/Sound Levels:
The user has two options for preset alarm levels: FS209E and ISO standards. The
ISO is modified directly and the resulting FS209E is shown alongside. The user may
choose between the following preset alarm levels:
•
ISO 3 / FS209E 1
•
ISO 4 / FS209E 10
•
ISO 5 / FS209E 100
•
ISO 6 / FS209E 1,000
•
ISO 7 / FS209E 10,000
•
ISO 8 / FS209E 100,000
•
ISO 9 / FS209E 1,000,000
The custom alarm level is a count limit imposed on every channel of the HPC301. If
any channel were to exceed this threshold, the alarm will sound.
•
If the PRESET alarm is enabled, the channel sizes of the HPC301 will be
locked to 0.3 μm 0.5 μm and 5.0 μm. This is because the ISO and FS209E
standards used in the HPC301 are limitations set for these channel sizes.
•
Enabling the Custom alarm does not inhibit the user from changing channel
sizes.
The sound option maybe be set to ON or OFF. Setting this option to OFF will not