background image

                                                                 

                 

                   

 

Page 23 of 33 

 

V1.1 

affect the alarm.   

 

Print/Save Options: 

A Bluetooth enabled printer is required for printing; contact Hal Tech for a quote. The 

Print option can be set to either ON or OFF. When set to ON, the unit will print results 

after a measurement based on the setting of the Save option. If the save option is set 

to ON or OFF and the print option is ON, then every measurement will be printed. If 

the  save  option  is  set  to  Average  and  the  print  option  is  on,  only  the  average 

measurement will be printed at the end of the measurement process.   

 

The save option can be set to one of three options: ON, OFF, and Average. If set to 

Off, no data is saved. If set to On, every sample is saved. If set to Average, only the 

average of all the measurements is saved.   

 

 

The optional Bluetooth printer is required to print. If you would like to purchase 

the printer, contact Hal Technology (

[email protected]

). 

 

Bluetooth Settings:   

The Bluetooth settings are for the optional Bluetooth enabled printer, which would be 

included with the HPC301 if purchased. If you do not use the printing feature of this 

unit, be sure to set the Bluetooth Power option to OFF. This will decrease the drain 

on the battery. The Address option can be modified, and will correspond to the 6-byte 

(48-bit)  Bluetooth  address  of  the  printer.  If  the  printer  was  purchased  from  Hal 

Technology when the HPC301 was ordered, the printer will already be configured. 

Selecting Connect and pressing the 

ENTER

 key will test the Bluetooth connection 

between  the  module  and  the  printer.  Be  sure  that  the  printer  is  turned  on  before 

attempting the Connect test.     

 

Factory Settings:   

The firmware  update  option  is reserved for future use  by  Hal Technology,  and the 

Содержание HAL-HPC301

Страница 1: ...f 33 V1 1 HAL HPC301 Handheld Particle Counter Operational Manual 7970 Cherry Ave Suite 303 Rancho Cucamonga CA 91730 USA Phone 866 438 4258 Fax 855 402 9190 Info haltechnologies com http haltechnolog...

Страница 2: ...Page 2 of 33 V1 1 This page intentionally left blank...

Страница 3: ...tion 12 2 1 Handheld Unit Overview 12 2 2 Key Functions 12 2 3 Graphical User Interface GUI 13 2 3 1 Main Screen 13 2 3 2 Measure Screen 14 2 3 3 Browse Screen 16 2 3 4 Settings Screen 17 2 3 5 Modify...

Страница 4: ...33 V1 1 Version Table Date Version Comment February 10 2017 1 1 Added Data Downloader instructions December 20 2016 1 0 Formal Release April 15 2015 0 2 Pre release updating April 7 2015 0 1 Pre relea...

Страница 5: ...oducts it describes at any time without notice or obligation Published in the United States of America Copyright 2016 by Hal Technology All rights reserved No part of the contents of this manual may b...

Страница 6: ...hat may be damaged by improper handling The warranty is void for any unauthorized opening of the instrument Environmental Requirements To avoid accidents or damage to the instrument please avoid using...

Страница 7: ...33 V1 1 annually at least Please contact Hal Technology to schedule your calibration or any services needed The HAL HPC301 can only be serviced at Hal Technology or by a Hal Technology authorized pro...

Страница 8: ...light and generate electrical pulses proportional to the size of the incident particles The HAL HPC301 is a completely self contained particle counter that is no other parts are required to start moni...

Страница 9: ...ons are available to include Mass Concentration and Particulate Matter measurements Optional accessories also include a digital temperature relative humidity and pressure probe as well as an isokineti...

Страница 10: ...accurate measurement Up to 260 unique location labels No less than 6 hours of continuous operation Bluetooth Wireless printing capabilities Compatible with Window XP 7 8 10 operation systems 1 2 Spec...

Страница 11: ...er battery discharged Interface USB Bluetooth Power Rechargeable Lithium ion battery 7 4V 5200mAh AC adapter 100 240VAC to 9VDC 1 5A Dimension 93 W 180 H 48 D mm Weight Approximately 900 grams Environ...

Страница 12: ...rface is composed of a six key interface ENTER BACK Each button controls the following features Power button While the unit is off press and hold the power button for approximately 2 seconds to turn o...

Страница 13: ...flashes during charge cycle and becomes steady after the charge cycle is finished Ethernet Interface This feature is not standard for the HPC301 and is a future option 2 3 Graphical User Interface GUI...

Страница 14: ...reen one can navigate to the Settings Screen by pressing the UP key the Browse Screen by pressing the DOWN key or start taking data by pressing the RUN STOP key The Measure Screen also shows the statu...

Страница 15: ...al is to be set below 5 seconds the other must be greater or equal to 5 seconds For example if you want a Sample Duration of 4 seconds or less the Sample Interval must be at least 5 seconds The instru...

Страница 16: ...easurements can be calculated as 2 3 3 Browse Screen Figure 6 shows the Browse Screen of the HPC301 From this screen one can navigate to the Measure Screen by pressing the UP key the Settings Screen b...

Страница 17: ...From this screen one can navigate to the Browse Screen by pressing the UP key the Measure Screen by pressing the DOWN key or enter the settings screen by pressing the ENTER key All the unit s paramet...

Страница 18: ...e editing 00 00 The UP and DOWN keys will now edit the Sample Duration appropriately When finished press the BACK key to stop editing 00 00 Particle Sizes Modify particle sizes on Channels1 6 Sample P...

Страница 19: ...low the user to modify the following parameters over the following ranges Start Delay 00 00 to 59 59 Sample Interval 00 00 to 59 59 Sample Duration 00 00 to 59 59 Number of Samples 0 to 500 Material P...

Страница 20: ...et to Avg only the average across the number of samples will be stored When setting this number keep in mind the run time of the entire sampling procedure Depending upon the battery charge the user ma...

Страница 21: ...m the following standard Count Modes Cumulative Counts Differential Counts Counts Per Liter Counts Per Cubic Foot Counts Per Cubic Meter and the following optional Count modes Mass Concentration Parti...

Страница 22: ...he following preset alarm levels ISO 3 FS209E 1 ISO 4 FS209E 10 ISO 5 FS209E 100 ISO 6 FS209E 1 000 ISO 7 FS209E 10 000 ISO 8 FS209E 100 000 ISO 9 FS209E 1 000 000 The custom alarm level is a count li...

Страница 23: ...ed The optional Bluetooth printer is required to print If you would like to purchase the printer contact Hal Technology info haltechnologies com Bluetooth Settings The Bluetooth settings are for the o...

Страница 24: ...uired contact services haltechnologies com The Clear Last Record option is available in case the unit were to lose power while saving data If this were to occur it is possible the last record could be...

Страница 25: ...the instructions in the file named Readme txt that is usually provided on the compact disc After successful installation it is preferable to have the HPC301 powered off when using the software to dow...

Страница 26: ...d Go to the Factory Settings page under Settings and Clear the last record If this does not fix the problem try a Factory Reset all memory will be cleared The unit does not appear to be saving my chan...

Страница 27: ...of 33 V1 1 Printer does not print 1 Ensure the printer is in working order turned on and has thermal paper 2 Ensure Bluetooth is enabled on the HPC301 3 Run a connection test from the Bluetooth Settin...

Страница 28: ...un specified environmental condition Abnormal operation due to required instrument calibration Limitation of Warranty A Hal Technology warrants that all equipment shall be free from defects in materia...

Страница 29: ...Hal Technology s option be terminated C Warranty repairs shall be completed at a Hal Technology authorized service location by an authorized service technician or on site at buyer s facility by a Hal...

Страница 30: ...he warranty period Buyers may select a faster method of shipment at their own expense Contact Information HAL TECHNOLOGY LLC 7970 Cherry Avenue Suite 303 Rancho Cucamonga CA 92336 USA Phone 866 438 42...

Страница 31: ...a CA 92336 USA Fax 855 402 9190 toll free Or send relevant registration information to the email address below services haltechnologies com Company Contact Person Address City State Country Postal Zip...

Страница 32: ...o meets the 0 3 m threshold So both the 0 3 m and 2 5 m channel counts are incremented Differential Counts The relative number of counts for a channel referenced to the next largest channel For exampl...

Страница 33: ...Page 33 of 33 V1 1 Time Passing Wait Sample Wait Sample Wait Sample Wait Sample...

Отзывы: