Note:

 To calibrate the bargraph you must be able to input two input signals. Usually the mini-

mum input (LO Input) and the maximum input  (HI Input) signals are used for optimum accuracy. 

However a scaling can be accomplished with any two signals that are higher or lower than each 

other by more than 5% of fullscale and are not greater than +2.1VDC or less than -1.05VDC.

STEP C  SET THE LOW INPUT SIGNAL READING ON THE BAR

 

1)  Apply the LO input signal (4 mA in this example) to the input pins.

 

2)  Using the ZERO button adjust the bar down to the required position. 

STEP D  SET THE HIGH INPUT SIGNAL READING ON THE BAR

 

1)  Apply the high input signal (20mA in this example) to the input pins.  

Using the SPAN button adjust the bar to the required position.  

This position could be higher or lower than the position adjusted in  

Step 2. The scaling for an input of 4 to 20mA is now complete.

Apply 4 mA 

to the 

Input Pins 

and adjust

bar display 

to the required 

position

100

0

SP1 SP2 SP3 SP4

Zero Span

Apply 20 mA 

to the

Input Pins 

and adjust

bar display 

to the required 

position

100

0

SP1 SP2 SP3 SP4

Zero Span

STEP A  SELECT THE CENTER ZERO DISPLAY MODE FOR CH1

 

1)  Following the instructions on page 4, remove the meter from the case.

 

2)  Select the Center Zero Mode for CH1 by repositioning the jumper clip on the 

Center Zero Display Mode Select Header.

STEP B  REVIEW THE INPUT MODULE STATUS

 

1)  See pages 15 – 21 for information on input modules that may be use with this meter.

 

2)  Only the IDP4 Universal Input module can be used for dual inputs and information 

on this module can be found on page 15.

 

3)  Confirm that the correct range and input is selected on the input signal conditioning 

module.

Note:

 When undertaking an initial set up and primary scaling and calibration of the meter it is 

best to start with a reset of the scaling.

STEP C  RESET THE SCALING ON CHANNEL ONE

 

1)  Apply power to the meter and press the CH1 ZERO and CH1 SPAN buttons  

simultaneously for 2 seconds. This erases any previously memorized scalings,  

and resets the scaling to the factory default, of approximately zero to full scale,  

for an input, that is 0 to 100% of the range selected on the input signal  

conditioner.

The procedure for scaling the bar graph for bipolar signals is very simple. If say CH1 has to be scaled for -1V to +1V, the steps 

are as follows:

CH1

CH2

CH1, Center 

Zero Mode 

Selected

100

0

SP1 SP2 SP3 SP4

Zero Span

Reset the scaling 

to the default value

on by pressing

the Zero and

Span buttons

simultaneously

for 2 secs.

Standard Display Mode Calibration Procedure Continued

One Point Quickset Rescaling and Calibration Procedure

Center Zero Mode Scaling For Bipolar Inputs

ONE POINT RECALIBRATION

As explained earlier, the TFL-B101Q bargraph is calibrated using two point calibration. Once a bargraph is calibrated, the 

low end of the range may be then recalibrated without affecting the calibration of the high end, and 

vice versa.
For example, take an TFL-B101Q that has been calibrated to read zero to full scale for an input of 4 to 

20mA. If now the scaling has to be changed to read zero to full scale for an input of 0 to 20mA, only 

the low (4 mA) end needs to be recalibrated. The high (20 mA) end of the scaling is left untouched, 

and so does not change. The following one point recalibration procedure is used for this purpose.

STEP A  RECALIBRATE THE LOW INPUT SIGNAL READING ON THE BAR

 

1)  Apply the LO input signal (0 mA in this example) to the input pins.  

The first segment will flash, indicating an under range condition.

 

2)  Using the ZERO button adjust the bar up to the required position.

 

3)  The TFL-B101Q has now been recalibrated to read zero to fullscale  

for a 0 to 20 mA input.

Apply 0 mA 

to the Input Pins 

and adjust

bar display 

to the required 

position

100

0

SP1 SP2 SP3 SP4

Zero Span

Содержание TFL-B101Q

Страница 1: ...ly AC DC Auto sensing wide range supply PS1 std 85 265 VAC 95 370 VDC 2 5W max 4 2W PS2 15 48 VAC 10 72 VDC 2 5W max 4 2W Operating Temp 0 to 60 C Storage Temp 20 C to 70 C Relative Humidity 95 non co...

Страница 2: ...dia grams are used to visually assist in following programming steps The following symbols are used to represent the functions and displays of the meter When two fingers are shown side by side the tw...

Страница 3: ...side the bar display area and by an OFF segment inside the bar display area Horizontal or Reverse Mounting Meters can be mounted horizontally in the panel and for those applica tions that require an o...

Страница 4: ...driver to release two plastic catches on either side of the case and levering backwards Step 4 Starting with the top first insert a small flat bladed screw driver and carefully lever the plastic catc...

Страница 5: ...the Scaling by pressing the ZERO and SPAN buttons simultaneously for two seconds Both scaling memories will be erased and an internal default scale factor will be loaded This provides a display of zer...

Страница 6: ...libration of the meter it is best to start with a reset of the scaling STEP C RESET THE SCALING ON CHANNEL ONE 1 Apply power to the meter and press the CH1 ZERO and CH1 SPAN buttons simultaneously for...

Страница 7: ...e Display Driver Board See the layout diagram on Page 4 and 14 for the exact location The steps to setup the setpoints are as follows 1 SELECT THE RELAY ACTIVATION MODE FOR EACH INSTALLED RELAY Make s...

Страница 8: ...ity as to the signal status especially just after transitioning to a new zone However if two or more setpoints with differently specified colors are positioned at the same set point value the color sp...

Страница 9: ...t the analog output as measured on the external multimeter to be the required value 20mA in this example When the HI button is pressed the UP or DOWN indicator LED shows the direction of change Releas...

Страница 10: ...s 23 and 24 have a unique plug and socket outline to prevent cross connection The main board uses standard right angled connectors Replacement 2 3 and 4 pin plug connectors are available see Accessori...

Страница 11: ...EXC Display Board Front SP1 SP2 SP3 SP4 Zero Span 0 10 20 30 40 50 60 70 80 90 100 Display Board Front SP1 SP2 SP3 SP4 Zero Span 0 10 20 30 40 50 60 70 80 90 100 Zero SP4 SP3 SP2 SP1 100 0 10 30 50 70...

Страница 12: ...ated I Series input modules that have span or zero potentiometers can be interchanged between any I Series com patible meter without recalibration because all of the analog scaling and reference circu...

Страница 13: ...fset 0 _ Increase Span Decrease 141E I Series Input Signal Conditioning Modules IF02 Line Frequency SPAN ZERO Frequency to Voltage Conversion 60 to 500VAC 269D IA12 AC Millivolt RMS Sigma Delta 371A A...

Страница 14: ...re IS02 Pressure Load Cell 5 10VDC Exc 20 2mV V 4 6 wire 2 20 mV V 10V 5V 4W 6W EXC Pressure Transducer PRESSURE 151A IR03 Linear Potentiometer 1K min POTENTIOMETER 1K Minimum Digital Scaling Input Ex...

Страница 15: ...tting Right angled Connector 6mm 0 24 10mm 0 39 26 5mm 1 05 Push On Connector for FI series IO board The adapter snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an...

Страница 16: ...ader position 4 3 2 1 1 2 3 4 5 75 Turn Potentiometer 0 Equivalent Circuit Increase Zero Decrease 5 4 3 2 1 Decrease Zero Increase 1 2 3 4 5 75 Turn Potentiometer 0 Zero Pot Disabled SPAN Turn Clockwi...

Страница 17: ...have one 10 Amp relay or two 5 Amp relays plus 16 bit analog output or an extra 5 Amp relay Auto sensing AC DC wide range power supplies 85 265 V AC 95 370 V DC or 18 36 V AC 9 60 V DC Quick easy mou...

Страница 18: ...HAKARU PLUS CORPORATION Tel 81 6 6300 2195 Page 18 05 12 16 TFL B101Q DS FL6 Space...

Страница 19: ...40 40 30 35 25 20 0 5 10 15 9x5 45 45 40 30 35 25 20 0 5 10 15 10x5 50 50 45 40 30 35 25 20 0 5 10 15 12 15 0 3 6 9 1 10 100 1000 1 5 15 150 1500 2 20 200 2000 2 5 25 250 2500 3 30 300 3000 4 40 400 4...

Страница 20: ...e IS05 Pressure Load Cell 20 2mV V 5 10V Exc 4 wire IS06 Pressure Load Cell Ext Exc 20 2mV V 4 wire IS07 Pressure 20 2mV V with High Impedance and External Excitation IT03 RTD 100 Pt 2 3 4 wire 200 to...

Отзывы: