26

Optional Accessories

When using the temperature recorder, please refer to the “User manual for Temperature Recorder” 
provided with the recorder.

For installation and operation of the backup cooling system, please refer to the user's manual 
provided with the backup cooling system.
If the free-standing backup system is to be connected to DW-86L490J/490JA and DW-86W100J, it’s 
necessary to buy the relative switch box and plate.

The temperature recorder should only be installed by professionals or Haier 
serviceman.

Please purchase steel cylinders fi tted with a siphon, the siphon can be used to siphon 
up fl uids that fl ow out from the bottom of the steel cylinders.

Before installing the temperature recorder, please cut off  the power supply to avoid 
electric shock or fi re.

• The installation place of any equipment using CO

/LN

2

 must be in a well ventilated 

area. Increased CO

2

 in the air may be harmful and even fatal. If the ventilation is poor, 

another method should be considered in order to ensure a safe working environment.
• If a CO

2

 / LN

2

 steel cylinder falls over or one of the valves is damaged, then the steel 

cylinder will be turned into an uncontrollable lethal projectile.
• The temperature of liquid CO

/ LN

2

 is extremely low, which could cause frostbite. 

When replacing the cylinder, please always wear a pair of protective glasses and 
protective clothes.

Caution

Caution

Warning

Warning

Temperature recorder

CO

2

 and LN

2

 backup cooling system

7

• An ULT freezer is usually sensitive to its operating environment. If a unit is not 
installed in the conditions mentioned above, it cannot operate reliably. Please improve 
the environmental conditions before using the equipment.
• It is prohibited to use the unit in an outdoor place. After the unit is rained on, there is 
a danger of electric shock.

For the equipment to achieve optimal operating conditions, an intended installation location should 
satisfy the following requirements.
• Do not install the unit in a confi ned place. The doorway should be large enough for the unit to freely 
enter or exit the room if necessary. This is to allow the unit to be repaired easily and timely to avoid 
damage to property.
• The location for installation should be fl at and fi rm.
• There should be good ventilation and no direct sunlight.
• The freezer unit cannot share the same power socket with other equipment. The power plug should 
be securely connected with the power socket.
• The power cord for the freezer should not be twisted or pinched.
• If the power cord needs to be extended. the cross-section of the extended cable must be no less 
than 2 mm² and no longer than 3 m.
• Before using the freezer, check the voltage supply. A voltage stabilizer to deliver rated voltage ±10% 
is recommended for areas where the voltage is known to be unstable. The voltage stabilizer should 
be rated at least 4000W.
• The freezer must be securely grounded.
• If the power socket is connected with a ground terminal, make sure to inspect it for proper 
connection before using the equipment.
• If the socket is not grounded, it must be connected to a grounded terminal by a qualifi ed technician.

• Do not ground the freezer through gas lines, water mains, telephone lines and 
lightning rods as this may lead to electric shock.
• After installation, the power plug must be easy to reach. In case of emergency, it is 
easy to unplug. Nothing should block the ventilation port of the freezer. 

Installation

Installation environment

Installation site

Caution

Warning

Ambient temperature:10

 to 32

.The ideal temperature is 18

 to 25

. If necessary, use an air-

conditioning system to achieve the required ambient condition.

Ambient humidity: less than 80%. At an environment of 32°C, humidity should be less than 57% RH.

The intended location should be low in dust count.

The intended location should be vibration and shock free.

The highest elevation the unit can work safely: 2000 m above sea level.

Input voltage: within Rated Voltage ±10 % .

• Overvoltage category: The transient state is Category II facility.
• Pollution Degree:2.

Содержание DW-86L338J

Страница 1: ......

Страница 2: ...26 Troubleshooting 28 Refrigeration and Wiring Diagram 29 Speci cations Technical data and Packing list 31 33 Packing List Model User Manual Instruction to install sapcers Plastic bag Ice Scraper Key...

Страница 3: ...mers Thanks for buying a Haier ULT Freezer to make better use of this manual and this product in order to prevent injuries to personnel and damage to the product Please read carefully and follow the d...

Страница 4: ...nicians approved by manufacturer Never install your ULT in an unprotected area If the unit is rained on there is a danger of electric shock Your Haier ULT must not be installed in a damp area or an ar...

Страница 5: ...environment or operator s health If the unit is not in use for a long period of time make sure the power cord is unplugged Deteriorated insulation of the power cord may lead to electric shock or re If...

Страница 6: ...he door to avoid pinching your hands Do not tilt the unit more than 45 degrees when moving the unit When moving the unit please be careful not to stumble with the unit which could cause injury to pers...

Страница 7: ...zer over time Two access ports are installed in the back wall of the unit which can be used as the through hole for the thermocouple wires during testing and validation After all test wires are let th...

Страница 8: ...erature computer control to prevent the normal redundant systems from being erroneously controlled by the low temperature compressor 4 User friendly Design Equipped with LED display which can show the...

Страница 9: ...sfy the following requirements Do not install the unit in a con ned place The doorway should be large enough for the unit to freely enter or exit the room if necessary This is to allow the unit to be...

Страница 10: ...on for inspection or properly discard the battery Recycling the Rechargeable Battery Location of the battery The battery in the electric cabinet is for the power outage alarm lt is located inside the...

Страница 11: ...ay not work properly To do so please contact an Haier Equipment and Instrument Service Center Disposal of the freezer If the unit is to be stored unused in an unsupervised area for an extended period...

Страница 12: ...stem If the unit is not in use for a long period of time make sure the power cord is unplugged Deteriorated insulation of the power cord may lead to electric shock or re If the freezer is not in use i...

Страница 13: ...than 82 the set point temperature will automatically default to 82 If the ambient temperature is equal to or cooler than 30 the set point will resume at the intended inner set point This feature exte...

Страница 14: ...n 32 and lower than 38 Alarm indicator ashes Ambient temperature exceeds 38 Alarm indicator ashes Buzzer sounds intermittently probe failure alarm Main cabinet temperature control sensor fails Alarm i...

Страница 15: ...nditions Locked indicator on indicates all settings are locked to prevent erroneous operations Network indicator on indicates the network is in working mode Run indicator on indicates the compressors...

Страница 16: ...will be saved and the settings will be automatically exited from Then the temperature display will normally show the inner temperature Push for 3 seconds Push for 5 seconds During the above procedure...

Страница 17: ...tands for Years Months Days Hours Minutes respectively 2 Press Set key the temperature display shows years and ashes default value 10 Press or key select years The choices are from 10 to 99 If you wan...

Страница 18: ...the value set have been input into the computer Otherwise the setting is invalid For example If inner temperature is set to 80 setting the high temperature alarm to 75 is recommended Setting the high...

Отзывы: