Haier DW-86L338J Скачать руководство пользователя страница 13

22

• A fl ashing alarm cannot be cancelled unless the malfunction is eliminated. The 
buzzing alarm can be temporarily silenced for 30 minutes by pressing the "Silence" key. 
However, If the problem is not fi xed. the buzzer alarm will resume after 30 minutes.
• When using the freezer, the battery switch must be turned on to charge the battery.
• When there is a power outage, the battery sustains the temperature display. If the 
battery voltage is insuffi  cient. the temperature display will turn off .
• While the battery is still capable of providing power to the display, the temperature 
display can be turned off  by unplugging the power cord and turning off  the battery 
control switch.
• The freezer is also designed to auto-adjust the inner temperature set point in high 
ambient temperatures When the ambient temperature is warmer than 35

 and the 

set point temperature is set to be lower than -82

, the set point temperature will 

automatically default to -82

. If the ambient temperature is equal to or cooler than 

30

,the set point will resume at the intended inner set point. This feature extends the 

life expectancy of the freezer.

Caution

Setting the buzzer alarm resumption time

• When the unit is in the alarm mode, you may press the “Silence” key to stop the buzzing of the 
alarm (The remote alarm cannot be cancelled).
• If the alarm condition still persists, the buzzer alarm will resume automatically after 30 minutes.

Remote alarm terminals

• Remote alarm terminals are located on the control 
box on the right side of the unit. The alarm signals 
are delivered via the terminals. The terminals are 
rated for 30V DC at 2A.
• Terminal output:
Remote alarm terminals consists of NC, NO and 
COM terminals. Users can choose the normal Open 
or Normal Close alarms as needed.

 REMOTE ALARM  

Optional

 MAX   DC30V  2A

Battery Switch

POWER

on

off

on

off

SWITCH

11

Freezer parts and control panels

Freezer parts

•DW-86L338J/338JA/388J/486E/578J/578JA/628E/728J/728JA/828J/828JA

Содержание DW-86L338J

Страница 1: ......

Страница 2: ...26 Troubleshooting 28 Refrigeration and Wiring Diagram 29 Speci cations Technical data and Packing list 31 33 Packing List Model User Manual Instruction to install sapcers Plastic bag Ice Scraper Key...

Страница 3: ...mers Thanks for buying a Haier ULT Freezer to make better use of this manual and this product in order to prevent injuries to personnel and damage to the product Please read carefully and follow the d...

Страница 4: ...nicians approved by manufacturer Never install your ULT in an unprotected area If the unit is rained on there is a danger of electric shock Your Haier ULT must not be installed in a damp area or an ar...

Страница 5: ...environment or operator s health If the unit is not in use for a long period of time make sure the power cord is unplugged Deteriorated insulation of the power cord may lead to electric shock or re If...

Страница 6: ...he door to avoid pinching your hands Do not tilt the unit more than 45 degrees when moving the unit When moving the unit please be careful not to stumble with the unit which could cause injury to pers...

Страница 7: ...zer over time Two access ports are installed in the back wall of the unit which can be used as the through hole for the thermocouple wires during testing and validation After all test wires are let th...

Страница 8: ...erature computer control to prevent the normal redundant systems from being erroneously controlled by the low temperature compressor 4 User friendly Design Equipped with LED display which can show the...

Страница 9: ...sfy the following requirements Do not install the unit in a con ned place The doorway should be large enough for the unit to freely enter or exit the room if necessary This is to allow the unit to be...

Страница 10: ...on for inspection or properly discard the battery Recycling the Rechargeable Battery Location of the battery The battery in the electric cabinet is for the power outage alarm lt is located inside the...

Страница 11: ...ay not work properly To do so please contact an Haier Equipment and Instrument Service Center Disposal of the freezer If the unit is to be stored unused in an unsupervised area for an extended period...

Страница 12: ...stem If the unit is not in use for a long period of time make sure the power cord is unplugged Deteriorated insulation of the power cord may lead to electric shock or re If the freezer is not in use i...

Страница 13: ...than 82 the set point temperature will automatically default to 82 If the ambient temperature is equal to or cooler than 30 the set point will resume at the intended inner set point This feature exte...

Страница 14: ...n 32 and lower than 38 Alarm indicator ashes Ambient temperature exceeds 38 Alarm indicator ashes Buzzer sounds intermittently probe failure alarm Main cabinet temperature control sensor fails Alarm i...

Страница 15: ...nditions Locked indicator on indicates all settings are locked to prevent erroneous operations Network indicator on indicates the network is in working mode Run indicator on indicates the compressors...

Страница 16: ...will be saved and the settings will be automatically exited from Then the temperature display will normally show the inner temperature Push for 3 seconds Push for 5 seconds During the above procedure...

Страница 17: ...tands for Years Months Days Hours Minutes respectively 2 Press Set key the temperature display shows years and ashes default value 10 Press or key select years The choices are from 10 to 99 If you wan...

Страница 18: ...the value set have been input into the computer Otherwise the setting is invalid For example If inner temperature is set to 80 setting the high temperature alarm to 75 is recommended Setting the high...