• If the inside temperature arises because of the failure of the freezer, which cannot be solved 
within short time, please remove the sample to avoid the potential damage.

• Before putting samples in the freezer to be stored, fi rst check that the freezer's temperature for 
the samples conforms with the temperature that is required for the samples, in order to prevent 
the samples to be stored from getting damaged or lost due to the freezer not attaining the 
temperature required.

• Because it takes time for the refrigeration temperature to reach the stored samples, there 
is normally a temperature discrepancy between the actual shown temperature and the set 
point. This is a normal phenomenon. The lower the set point is, the smaller this temperature 
discrepancy becomes.

•All ultra low temperature storage units are low temperature storage equipment. It is prohibited 
to load an excessive amount of samples into the unit at one time. The compressors run 
for a long period of time without stopping. The freezer temperature may not decrease, and 
the compressors can become overheated. Samples must be loaded in batches, and while 
incrementally decreasing the temperature setting. The process should be repeated until the fi nal 
temperature is reached.

• Do not use any unauthorized mechanical tools or other means to accelerate the defrosting 
process.

• Do not damage the refrigeration circuit.

• Do not use any non manufacturer-approved electrical components in the freezer.

The Haier ULT freezer control setting is stored in its memory system. Should there be a power outage 
and recovery, the unit can resume its operation based on the previous settings.

• In the event of a power outage and recovery, be sure to wait for at least 5 minutes 
before turning the unit on again to avoid damaging the compressors and refrigeration 
system.
• If the unit is not in use for a long period of time, make sure the power cord is 
unplugged. Deteriorated insulation of the power cord may lead to electric shock or fi re.
• If the freezer is not in use in an area without any supervision, please make sure 
children will not approach the freezer and the doors should not be closed.

Operation after a Power Outage

Caution

23

• To prevent electric shock or injury to operators, the AC power supply to the freezer 
must be disconnected completely before any repair and maintenance work is to be 
performed.

• During any repair maintenance work, do not inhale medical particles or aerosols near 
the equipment as they might be harmful to your health.

Cleaning maintenance

Warning

Cleaning the freezer

• Clean the unit once a month.This can help the exterior look new.

• Use a dry cloth to wipe away loose dust inside and outside of the freezer. If the unit is rather dirty, 
use a clean doth soaked with a neutral detergent to clean the unit. Then use a dry cloth to wipe away 
any residual detergent solution.

• Never pour water onto or into the unit. Doing so can damage the electric insulation and cause 
failure.

• Compressors and other mechanical parts are hermetically sealed. They do not need lubrication.

• The users can easily remove the frost or ice on the chamber and clean the condensaor fi lter as 
often as necessary.

Cleaning the freezer

Clean the condensor fi lter when the control panel shows an signal for "Hot Condensor" and the alarm 
fl ashes. Even if the light is not on, the condensor fi lter should be checked regularly according to the 
suggestion from the distributor.

To clean the fi lter, follow the procedure below.
1. Pull off  the front grill cover.
2. Pull out the fi lter screen.
3. Use water to wash the fi lter screen.
4. After the fi lter screen is dry, reinstall it back
in its original position and close the cover.
5. If the “Hot Condenser” light is on before 
cleaning, check the light to make sure that it shuts 
off  after cleaning. If it does not shut off , please 
contact after-sales service personnel.

Содержание DW-86L338J

Страница 1: ......

Страница 2: ...26 Troubleshooting 28 Refrigeration and Wiring Diagram 29 Speci cations Technical data and Packing list 31 33 Packing List Model User Manual Instruction to install sapcers Plastic bag Ice Scraper Key...

Страница 3: ...mers Thanks for buying a Haier ULT Freezer to make better use of this manual and this product in order to prevent injuries to personnel and damage to the product Please read carefully and follow the d...

Страница 4: ...nicians approved by manufacturer Never install your ULT in an unprotected area If the unit is rained on there is a danger of electric shock Your Haier ULT must not be installed in a damp area or an ar...

Страница 5: ...environment or operator s health If the unit is not in use for a long period of time make sure the power cord is unplugged Deteriorated insulation of the power cord may lead to electric shock or re If...

Страница 6: ...he door to avoid pinching your hands Do not tilt the unit more than 45 degrees when moving the unit When moving the unit please be careful not to stumble with the unit which could cause injury to pers...

Страница 7: ...zer over time Two access ports are installed in the back wall of the unit which can be used as the through hole for the thermocouple wires during testing and validation After all test wires are let th...

Страница 8: ...erature computer control to prevent the normal redundant systems from being erroneously controlled by the low temperature compressor 4 User friendly Design Equipped with LED display which can show the...

Страница 9: ...sfy the following requirements Do not install the unit in a con ned place The doorway should be large enough for the unit to freely enter or exit the room if necessary This is to allow the unit to be...

Страница 10: ...on for inspection or properly discard the battery Recycling the Rechargeable Battery Location of the battery The battery in the electric cabinet is for the power outage alarm lt is located inside the...

Страница 11: ...ay not work properly To do so please contact an Haier Equipment and Instrument Service Center Disposal of the freezer If the unit is to be stored unused in an unsupervised area for an extended period...

Страница 12: ...stem If the unit is not in use for a long period of time make sure the power cord is unplugged Deteriorated insulation of the power cord may lead to electric shock or re If the freezer is not in use i...

Страница 13: ...than 82 the set point temperature will automatically default to 82 If the ambient temperature is equal to or cooler than 30 the set point will resume at the intended inner set point This feature exte...

Страница 14: ...n 32 and lower than 38 Alarm indicator ashes Ambient temperature exceeds 38 Alarm indicator ashes Buzzer sounds intermittently probe failure alarm Main cabinet temperature control sensor fails Alarm i...

Страница 15: ...nditions Locked indicator on indicates all settings are locked to prevent erroneous operations Network indicator on indicates the network is in working mode Run indicator on indicates the compressors...

Страница 16: ...will be saved and the settings will be automatically exited from Then the temperature display will normally show the inner temperature Push for 3 seconds Push for 5 seconds During the above procedure...

Страница 17: ...tands for Years Months Days Hours Minutes respectively 2 Press Set key the temperature display shows years and ashes default value 10 Press or key select years The choices are from 10 to 99 If you wan...

Страница 18: ...the value set have been input into the computer Otherwise the setting is invalid For example If inner temperature is set to 80 setting the high temperature alarm to 75 is recommended Setting the high...

Отзывы: