Chapter 7, Troubleshooting and Maintenance
87
Haemonetics
®
TEG
®
6s User Manual
P/N 115191-IE, Manual revision: AB
3019
Firmware update failed
There is a mismatch
between the existing
firmware and the updated
firmware, which causes
the automatic update to
fail.
Turn the power off and
then back on to reboot the
system. If the problem
persists, update the
firmware again or reset
the system to the
previous firmware
revision.
Note: This action
requires
administrator
access; if no
administrator is
available or if the
problem persists,
contact the local
Haemonetics
representative.
3027
Power-On-Self Test Fault -
Diskspace is critically low
The Power-On Self Test
(POST) detected that the
available disk space is
below the minimum
threshold.
Delete unneeded test
results from the analyzer,
and then turn the power off
and back on to reboot the
system.
Note: This action
requires
administrator
access; if no
administrator is
available or if the
problem persists,
contact the local
Haemonetics
representative.
Error Code
LCD Display Message
Probable Cause
Recommended Action