Chapter 6, Quality Control
75
Haemonetics
®
TEG
®
6s User Manual
P/N 115191-IE, Manual revision: AB
10.Touch
results
to return to the results screen. When the test is complete,
the following changes occur on the results screen:
The test timer is replaced by the system date and time.
The status of each channel is shown on the right side of the screen. A
green check mark
indicates that QC passed for the channel, and
a red
indicates that QC failed for the channel.
If all four channels passed, the” QC overdue” icon at the top of the
screen no longer shows the QC level that passed.
The
stop
button is replaced by a
done
button.
The
button is enabled.
Figure 6-8, QC test results (complete)
11.When the analyzer displays the “Remove cartridge” prompt, pull the used
cartridge out of the slot and immediately dispose of it in a marked,
biohazard waste receptacle.
Note: A flashing light at the cartridge slot also indicates that it is safe to
remove a cartridge.