Inspection Procedures
3-7
P/N 108755-US, Manual revision: AA
T
ROUBLESHOOTING
Table 3-1, General Troubleshooting Guide
Failure
Diagnosis
Action
Overweight blood collec-
tion.
Shaker not clamping or not com-
pletely clamping (i.e., solenoid
throw requires adjusting).
See "Tube clamp not functioning" in
this table.
Shaker not calibrated.
Calibrate Shaker. See Calibration sec-
tion above.
Foreign object on shroud.
Remove object.
Underweight blood collec-
tion.
Shaker not calibrated.
Calibrate Shaker. See Calibration sec-
tion above.
Foreign object on shroud.
Remove object.
Shaker will not calibrate.
Faulty strain gauge.
See Chapter 5, Strain Gauge Inspec-
tion and Replacement on page 5-2.
Weight sensing assembly requires
adjustment.
Adjust R3 and R7 on body board.
Shroud "clunks," moves with
a jarring motion or fails to
move.
Something obstructing the shroud or
link.
Verify power cord is not blocking
shroud on its back stroke. Inspect area
around and under shroud for obstruct-
ing objects and remove them (with the
Shaker unplugged), if necessary.
Shroud boss broken loose from the
inside of the shroud.
Replace shroud. See Chapter 10,
Shroud Assembly Installation on page
10-8.
Shroud is rubbing base assembly or
link assembly.
Bow shroud. See Chapter 10, Bowing
the Shroud on page 10-7.
Set screws on crank assembly are
loose.
Replace screws.
Link M, Link Y or Shaft M damaged
or loose.
See Chapter 6, Link Assembly Inspec-
tion on page 6-2.
Shaker extremely noisy
(grinding or rubbing sound).
The motor and Link Y connections
have become worn or out of align-
ment.
See Chapter 6, Link Assembly Inspec-
tion on page 6-2.
Tube clamp fails to engage at
collection completion.
Insufficient power to solenoid.
See Chapter 7, Tube Clamp/Solenoid
Inspection, Replacement and Adjust-
ment on page 7-2.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...