Shroud and Base Assemblies
10-7
P/N 108755-US, Manual revision: AA
B
OWING
THE
S
HROUD
Shrouds shipped from the manufacturer may arrive with the sides slightly bowed
in from the pressure of packing material. Although this is not a permanent
condition, it should be corrected prior to installing the link assembly or the link
assembly may rub against the inside of the shroud and cause the Shaker to collect
overweight units of blood.
1.
Partially assemble the shroud by installing the bag pins, as described in
the previous section.
2.
Place the shroud in a sink or wash tub with the opening facing up. Insert
a 2 x 4 x 5.50 in. wooden block, padded on each end with 1 x 2 x 5 in.
foam pads, immediately behind and parallel to the front boss and wedge
it firmly in place.
3.
Fill the shroud with hot water 140
°
to 160
°
F (60
°
to 71
°
C) and allow the
shroud to soak for one hour. Pour off the water, leaving the block in place,
and allow the shroud to cool for at least 30 minutes.
4.
Remove the block and foam pads.
5.
Dry the shroud and prepare to install the link assembly.
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...