6-2
Link Assembly
P/N 108755-US, Manual revision: AA
L
INK
A
SSEMBLY
I
NSPECTION
Warning: Protective gloves should be worn while performing the following steps
and changed as often as necessary to prevent the transmission of biohazardous
materials to repair personnel and the surrounding area.
Caution:
Always wear an ESD strap when working on the Shaker.
1.
Remove the shroud (see Chapter 4, Shroud Removal on page 4-3). Posi-
tion the Shaker with the left side facing forward so the link is easily acces-
sible.
Figure 6-1, Link Assembly Inspection
2.
Inspect the .125" D x 4.750" L shaft connecting the link D and link Y to-
gether for security and ease of movement. The shaft should fit tightly into
the link Y at either end and should be peened in place at both ends. The
link M and link Y should move freely without excessive play. The Teflon
®
bearings should be free of any damage and should be secure in their
mounting holes. If damage or excessive play is noted, the entire assembly
must be replaced.
3.
Inspect the .125 D x 3.515 L in. shaft for damage and ease of movement.
Examine the Teflon bearings for any damage and security in their mount-
ing holes. The shaft should glide smoothly in and out of bearings without
excessive side to side play. If the shaft or bearings are damaged or have
excessive play, replace the parts necessary to return the Shaker to proper
operating condition.
Shaft
(.125 n D x 3.515 in. L)
Link D
Shaft
(.125 n D x 4.75 in. L)
Link M
Link Y
Содержание SEBRA 1040
Страница 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Страница 3: ......
Страница 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Страница 8: ...vi Preface P N 108755 US Manual revision AA...
Страница 12: ...x Table of Contents P N 108755 US Manual revision AA...
Страница 14: ...xii List of Figures P N 108755 US Manual revision AA...
Страница 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Страница 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Страница 37: ...P N 108755 US Manual revision AA Chapter 4 Parts Access BASE PLATE REMOVAL 4 2 SHROUD REMOVAL 4 3...
Страница 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Страница 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Страница 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Страница 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Страница 83: ...P N 108755 US Manual revision AA Chapter 11 Clamp Body Boss Replacement REPLACING THE BODY BOSS 11 2...
Страница 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Страница 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Страница 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Страница 103: ...P N 108755 US Manual revision AA Chapter 15 Shaker Keystroke Quick Reference KEYSTROKE QUICK REFERENCE 15 2...
Страница 105: ......
Страница 106: ...Haemonetics Corporation 400 Wood Road Braintree MA 02184 USA www haemonetics com 1995 2010 Haemonetics Corporation...