3-8
Describing the MCS+ Control Panel
P/N 85213-30, Manual revision: B
Pump Start/Stop key
This key can be used to either stop the pumps, or re-start the pumps (if stopped
by the operator). Pressing this key does not affect centrifuge operation.
Caution: If the pumps have been stopped using this key, and remain stopped for
longer than two to three minutes during DRAW later, the bowl may become over-
packed with red cells, because the centrifuge will continue to spin. This can cre-
ate a potential flow problem during RETURN. The operator should return the
bowl contents to the donor before proceeding with DRAW.
Cuff control key
During normal MCS+ operation, the cuff will automatically inflate during DRAW
and deflate during RETURN. The operator can use the cuff key to manually
control the pressure cuff prior to a procedure when performing the venipuncture,
or during READY or DRAW mode to modify cuff pressure. The cuff cannot be
inflated during RETURN.
Figure 3-7, MCS+ cuff control key
Programming
keys
Certain system operating parameters have been selected by Haemonetics as
default values. These parameters provide optimal results in MCS+ apheresis
procedures with the average donor, as well as for average collection require-
ments. However, it is possible to alter and subsequently retain the altered param-
eters for specific collection requirements using the MCS+ programming keys.
Figure 3-8, Example of MCS+ program parameters display screen
Note: Further explanation about these parameters is provided in each respective
MCS+ protocol manual.
Cuff Pressure
50
mmHg
Draw Speed
90 ml/min
Return Speed
120 ml/min
NaCl Vol/Cycle
0 ml
MODIFY PARAMETERS
Min Plt Weight
0 g
AC Ratio
1: 9
LDP
STOP
Cycle 0
0 Platelet Weight
0 ml Processed
Press MODIFY to select, +/- to change values.
Press SAVE to save all values.
Press HELP to return to Main display.
Содержание MCS+
Страница 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Страница 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Страница 20: ......
Страница 30: ......
Страница 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Страница 85: ......
Страница 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Страница 100: ......